Related Experiment Video
Updated: Jan 31, 2026

Assessing Early Stage Open-Angle Glaucoma in Patients by Isolated-Check Visual Evoked Potential
Published on: May 25, 2020
Clinical Outcomes of Two Versions of the Ahmed Glaucoma Valve FP7 in Glaucoma Patients
Styliani Alexia Papadonta1,2, Kalliopi Kontopoulou1, Elpida Kollia3,4,5
1Department of Ophthalmology, St. Johannes Hospital, Dortmund, DEU.
Background:
The primary goal of surgical treatment for glaucoma is the sustained reduction of intraocular pressure (IOP). The Ahmed® Glaucoma Valve (AGV) (New World Medical, Inc., Rancho Cucamonga, CA) is a surgically implanted drainage device that allows bidirectional aqueous humor outflow and IOP regulation. Although multiple generations of the AGV have been developed, comparative studies evaluating patient outcomes remain limited. This retrospective study aimed to compare clinical outcomes following implantation of two different generations of the AGV FP7 in patients with complicated, refractory glaucoma.
Materials And Methods:
This retrospective study analyzed preoperative and postoperative clinical data from glaucoma patients who underwent AGV FP7 implantation at the Eye Clinic of St. Johannes Hospital in Dortmund, Germany. Two cohorts were compared: group A, treated with an earlier AGV FP7 model between January 2008 and December 2012 (n = 159 eyes), and group B, treated with a newer AGV FP7 model between January 2019 and December 2020 (n = 109 eyes). The newer version features a highly polished surface designed to reduce friction and tissue reaction, whereas the earlier model has a comparatively rougher surface. Clinical outcomes were assessed over a one-year follow-up period and included IOP reduction, number of antiglaucoma medications, changes in visual acuity, postoperative complications, need for reoperation, and overall therapeutic success.
Results:
One year postoperatively, IOP reduction was significantly greater in group B (52.5%) compared to group A (42.4%). Despite this difference, the therapeutic target was achieved in both groups. The number of required topical antiglaucoma medications also decreased significantly following AGV implantation by 56.8% in group A and 68.6% in group B. The most common complication at discharge was ocular hypotony, observed in 30.8% of cases in group A and 19.6% in group B. However, no cases of persistent hypotony were reported in either group at the one-year follow-up.
Conclusion:
Both the earlier and newer versions of the AGV FP7 effectively provided sustained IOP reduction in patients with refractory glaucoma. Although the newer generation demonstrated a greater IOP reduction and was associated with fewer short-term complications, including hypotony, choroidal detachment, and endothelial touch, overall therapeutic success remained high in both groups.
Related Concept Videos
Glaucoma: Overview
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Angle Closure Glaucoma: Treatment
Predicting Reaction Outcomes
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Outcomes of Glycolysis
Cellular respiration can occur aerobically (with oxygen) or anaerobically (without oxygen). In the presence of oxygen, cellular respiration starts with glycolysis and continues with pyruvate...