Pacemaker Implantation Rates With the Self-Expandable Navitor Valve
Michael Paukovitsch1, Moritz Konjack1, Dominik Felbel1
1Department of Cardiology Ulm University Heart Center Albert-Einstein-Allee 23 89081 Ulm Germany.
Transcatheter aortic valve replacement with the Navitor self-expandable valve shows higher rates of permanent pacemaker implantation (PPI) compared to the Evolut system. High implantation depth may reduce PPI rates for Navitor, making outcomes comparable to other self-expandable valves.
Area of Science:
- Cardiology
- Interventional Cardiology
- Biomedical Engineering
Background:
- Self-expandable valves in transcatheter aortic valve replacement (TAVR) are associated with higher permanent pacemaker implantation (PPI) rates than balloon-expandable systems.
- The Navitor prosthesis is a novel, tubular-shaped self-expandable valve with increasing clinical use.
Purpose of the Study:
- To compare 30-day PPI rates between the Navitor and Evolut self-expandable TAVR systems.
- To investigate factors influencing PPI rates with the Navitor prosthesis.
Main Methods:
- A comparative study of 148 patients receiving the Navitor system and 165 patients receiving the Evolut system.
- Exclusion of patients with pre-existing PPI, high-risk ECG, or specific prosthesis sizes.
- Multivariable logistic regression analysis to adjust for baseline differences.
Main Results:
- The Navitor group was older and had a higher predicted mortality risk.
- Overall PPI rates were significantly higher with Navitor (17.6%) versus Evolut (6.7%), persisting after adjustment.
- In patients with small annuli, PPI rates were similar between the two systems.
- Higher PPI rates were observed with larger Navitor prostheses (27/29mm).
- Deeper implantation depth was associated with lower PPI rates in the Navitor cohort.
Conclusions:
- Navitor TAVR may have PPI rates comparable to other self-expandable valves, depending on prosthesis size.
- Optimizing implantation depth for the Navitor prosthesis could potentially reduce PPI rates.
More Related Videos
06:04Upper-extremity Approach for Secondary Access in Transfemoral Transcatheter Aortic Valve Implantation
Published on: August 8, 2025
05:31Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022
Related Concept Videos
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Reaction Rate
The mathematical representation of the change in the concentration of reactants and products, over time, is the rate...
Related Rates
Speciation Rates
Measuring Reaction Rates
Rate-Determining Steps
In a multistep reaction mechanism, one of the elementary steps progresses significantly slower than the others. This slowest step is called the rate-limiting step (or rate-determining step). A reaction cannot proceed faster than its slowest step, and hence, the rate-determining step limits the overall reaction rate.
The concept of rate-determining step can be understood from the analogy of a 4-lane freeway with a short-stretch of traffic-bottleneck caused due to...