Small field dosimetry on a ring-shape Linac machine with a dual-layer MLC
Antonella Fogliata1, Antonella Stravato2,3, Francesca Dusi4
1Humanitas Research Hospital IRCCS, Radiotherapy and Radiosurgery Dept, Milan-Rozzano, Italy.
Background:
The Varian Halcyon linac, equipped with a dual-layer MLC and a single 6 MV FFF beam, is increasingly used for treatments involving small target volumes. The accuracy of dose calculation for small static fields on this platform remains a matter of concern.
Purpose:
This study investigates the dosimetric accuracy of small static fields generated by the Halcyon dual-layer MLC and to evaluate their modeling in Eclipse through Acuros XB versions 16.1, 18.0, and 18.1.
Methods:
On-axis field out put factors (FOFs) were measured for square and rectangular fields (0.5-4 cm) according to TRS-483, using the equivalent square field formalism by Fogliata. Measurements with microDiamond and microSilicon detectors were compared with Acuros XB calculations in Eclipse versions 16.1, 18.0, and 18.1. Profiles and penumbrae were analyzed, and a clinical single-lesion brain VMAT plan was evaluated.
Results:
The two detectors agreed within ± 0.5%. Differences from vendor Representative Beam Data were 1.1%-1.8% for 1 cm fields. Acuros XB v.16.1 underestimated FOFs by up to -27% for 0.5 × 0.5 cm2, while v.18.x agreed within ∼2%. Y-direction penumbrae were larger for proximal bank or groove edges, not fully modeled in TPS. In the clinical case investigated, 50% of control points had fields < 1 cm, with central underdosage in v.16.1.
Conclusion:
Acuros XB ≥ 18.x improves small-field accuracy for Halcyon, although Y-penumbra modeling remains incomplete. Constraints on minimum field sizes during treatment optimization are important to ensure high accuracy in treatment delivery.
More Related Videos
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
08:35Electrotaxis Studies of Lung Cancer Cells using a Multichannel Dual-electric-field Microfluidic Chip
Published on: December 29, 2015
Related Concept Videos
Molecular Shape and Polarity
Machines
A free-body diagram of the...
VSEPR Theory and the Basic Shapes
Machines: Problem Solving II
Molecular Shapes
Two regions of electron density in a diatomic...
Machines: Problem Solving I
The toggle clamp system is a machine structure consisting of movable, pin-connected multi-force members that form a stabilized system to transmit forces. The...