Transient numerical simulation of hemodynamics in bioprosthetic heart valves: insights into valve sizing and
M Berger1, J Golks2, L Gleissner2
1Department of Medical Technologies, MCI - The Entrepreneurial School, Innsbruck, Austria. manuel.berger@mci.edu.
Purpose:
This study aims to evaluate the hemodynamic effect of tilting after implantation of a bioprosthetic heart valve in an enlarged aortic annulus using computational fluid dynamics (CFD). The objective is to define the cost of excessive enlargement in transprosthetical flow and identify the optimal prosthesis size and implantation angle to reduce thrombosis and early degeneration.
Methods:
Virtual implantation of a 23-mm bioprosthetic model was conducted in a prespecified virtual aortic annulus of 23 mm. CFD simulations were conducted to analyze the hemodynamic parameters, including wall shear stress and turbulent kinetic energy, for bioprosthetic heart valves of the 23 mm (perpendicular implantation), 25 mm, and 27 mm (tilted implantation at different angles) after virtual annular enlargement. The simulations utilized transient flow models and mesh convergence studies, to ensure numerical accuracy and clinical relevance.
Results:
The 23-mm valve implanted without annular enlargement and in an aligned fashion to the annulus exhibited higher wall shear stress (WSS) and shear stress (SS) values than the 25-mm valve implanted at 12° tilted position and better hemodynamics than the 27-mm valve at 25° tilted position. The 25-mm valve after annular enlargement and implantation at 12° tilted position achieved the best hemodynamic performance together with a 23% and 7% reduction of WSS and SS as compared to the 23-mm valve.
Conclusions:
Slight oversizing and tilting after annular enlargement yields the best performance of an aortic bioprosthetic valve with lowest WSS and balanced turbulence.
Related Concept Videos
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Mitral Valve Prolapse I: Introduction
Mitral Valve Prolapse II: Assessment and Management
Mitral Valve Prolapse III: Nursing Management
Anatomy of the Heart
Numerical Calculations
The solution to a problem is obtained using different methods. While manually solving algebraic symbols is one of the most common methods, the graphical method is often preferred. Computers...