Virtual simulation accurately predicts in vivo flow diverter length for intracranial aneurysm treatment: a
Maximilian Rothe1, Thomas Liebig2, Martin Renz3
1Institute of Neuroradiology, Technical University of Munich School of Medicine and Health, Munich, Germany maxi.rothe@t-online.de.
Background:
Accurate device sizing is crucial for successful flow diverter (FD) therapy in intracranial aneurysms. This study assesses the accuracy and clinical utility of the Ankyras virtual simulation software (Mentice AB, Gothenburg, Sweden) in predicting FD length across multiple device types and manufacturers.
Methods:
We retrospectively analyzed 193 FDs (7 device types) deployed in 180 patients with 230 intracranial aneurysms. Simulation-based prediction of effective lengths (simulated length (SL)) and nominal manufacturer specifications (labeled length) were compared with measured in vivo lengths (measured length (ML)). Performance was evaluated using simulation accuracy (SA), absolute error (AE), relative error (RE), length ratio, and correlation analysis.
Results:
Virtual simulation demonstrated good clinical usability, achieving a mean absolute deviation of only -1.38 mm at the proximal landing zone compared with ML. Simulation-based predictions showed significantly superior accuracy compared with nominal manufacturer specifications (SA 89.6±11.3% vs 81.8±13.7%, P<0.001), with mean AE reduced by 58% (-1.38 mm vs -3.34 mm) and mean RE by 42% (10.5% vs 18.2%). Strong correlations between SL and ML (r=0.900) validated predictive reliability across all tested device types. Centerline correction technology further enhanced parameters such as SA to 94.8±9.5%.
Conclusion:
Virtual simulation-based prediction of effective FD length using the Ankyras software showed reliable and clinically meaningful results, enabling accurate estimations of the proximal landing zone and overall length. Ankyras, along with comparable simulation platforms, may provide considerable potential to facilitate FD selection and implementation, especially in anatomically complex cases.
Related Concept Videos
Reliability and Validity
Aneurysm II: Clinical Manifestations and Diagnostic Studies
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and...
Virtual Work
In static equilibrium, a body can experience an imaginary or virtual movement, such as displacement or rotation. The virtual work done by a force is equal to the dot product of force and virtual displacement in the direction of the force. When it comes to virtually rotating a...
Predicting Molecular Geometry
Aneurysm I: Introduction