Can common household devices alter the settings of programmable shunt valves?
Jack Ramsay1, Georgios Bonanos2, Alvaro Villabona2
1School of Medicine, Newcastle University.
Objective:
Various household devices contain magnets or emit electromagnetic fields and it is hypothesised that the strength of these magnets may be enough to affect the settings on programmable cerebrospinal fluid shunt valves. The aim of this study is to identify changes to the valve settings of seven commonly used programmable valves, when they are near common household devices generating a magnetic field.
Methods:
Several commonly encountered household devices were used, with and without rotational movement, with reference to seven commonly used programmable valves (the Codman Certas and Hakim, Sophysa SPV and SM8, Medtronic Strata II Regular and Small, and the Miethke M.Blue Plus). The valve settings were checked before and after the interference with the household device, by using either the manufacturers tool kit for reading the settings or x-rays.
Results:
We demonstrated that the only valve which was not affected by any of the household devices, regardless of the distance or rotational movement, was the Miethke M. Blue Plus. In addition, the programmable shunts were more likely to experience change to their settings with rotational movement as opposed to no movement, and the two devices most likely to induce this change were a smart watch and kids toy magnets.
Conclusions:
Everyday household devices have the potential to affect the settings on programmable shunts, leading to concerns around the maintenance of shunt settings both in and out of hospital due to household and healthcare electromagnetic fields. Surgeons, and other healthcare professionals involved in the care of those with programmable CSF shunts, should be aware of the potential risks and utilise this information when making clinical decisions and advising patients.
More Related Videos
Related Concept Videos
Shunt Admittances
To model this effect, the method of images is employed. This method involves replacing the Earth with an image conductor that mirrors the original...
Common Ion Effect
Household Wiring And Electrical Safety
Alterations in Respiration II
In Biot's breathing, the respiratory rate and depth are irregular, alternating between periods of deep gasping and apnea. Common causes...
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Altered States of Awareness
The ingestion of substances like stimulants or hallucinogens leads to chemical alterations in the brain...