Related Experiment Video
Updated: Feb 4, 2026

Sterile Pericarditis in Aachener Minipigs As a Model for Atrial Myopathy and Atrial Fibrillation
Published on: September 24, 2021
Feasibility and reliability of a smartwatch to detect atrial fibrillation after cardiac surgery: a prospective study
Konrad Schreier1, Michael Borger2, Alireza Sepehri Shamloo1
1Department of Cardiac Electrophysiology, Heart Center Leipzig, Leipzig, Germany.
Background:
Atrial fibrillation, the world's predominant cardiac arrhythmia, frequently emerges as a complication post-cardiac surgery, leading to serious outcomes like strokes, heart failures, and even death. Due to its often-silent nature, detecting it can be challenging. Smartwatches present a potential solution, offering screening that is more rigorous.
Objective:
This prospective observational study sought to assess the Withings Scanwatch's efficacy in identifying postoperative atrial fibrillation.
Methods:
After cardiac surgery, patients received a Withings Scanwatch. Over a span of 24 h, both the smartwatch's photoplethysmography sensor and standard telemetry kept track of any atrial fibrillation incidents.
Results:
At the end of the study, data from 260 patients was available for assessment. Atrial fibrillation was identified in 32 of these patients, either via telemetry or the smartwatch. Our data revealed a sensitivity of 69.0%, specificity of 98.7%, a positive predictive value of 87.0%, and a negative predictive value of 96.2%.
Conclusions:
This clinical study is the first to evaluate the photoplethysmography sensor of the Withings Scanwatch, and it shows that the Scanwatch has high a specificity and moderate sensitivity in detecting postoperative atrial fibrillation. Thus, Scanwatch may support the conventional screening for atrial fibrillation, and potentially reducing complications and costs of atrial fibrillation. Because of lower than expected sensitivity this technology cannot replace conventional monitoring in postoperative patients.
Related Concept Videos
Amyloid Fibrils
Amyloid deposits were observed as early as 1639 in the liver and the spleen. In 1854, Rudolph Virchow performed iodine staining,...
Amyloid Fibrils
Reliability and Validity
Fibril-associated Collagen
For example, the type II collagen fibrils in cartilage have covalently bound type IX fibril-associated collagens at regular intervals. Other types of fibril-associated collagens are...
Blood Studies for Cardiovascular System I: Cardiac Biomarkers
The essential diagnostic tools for detecting myocardial necrosis and monitoring individuals suspected of having acute coronary syndrome (ACS) include:
Troponins
Troponins, particularly cardiac troponins I and T, are the most precise and sensitive markers of myocardial injury. They are detectable within 4-6 hours of myocardial injury and remain...
Distribution Reliability and Automation