When Stability Fails: Rare Very Late Dislodgement of 3830 Lead
Dilip Kumar1, Niladri Ghosh1, Ashesh Halder1
1Manipal Hospital EM Bypass Kolkata Kolkata India.
Clinical Case Reports
|February 2, 2026
Summary
Late dislodgement of a Medtronic SelectSecure lumenless lead occurred three years after successful left bundle branch area pacing. This rare event in pacemaker patients underscores the need for ongoing monitoring of cardiac pacing systems.
Area of Science:
- Cardiology
- Medical Device Technology
Background:
- Left bundle branch area pacing (LBBAP) is an advanced technique for physiological pacing.
- Lumenless leads offer specific advantages but may have unique complications.
- Very late lead dislodgement is an uncommon but significant complication.
Purpose of the Study:
- To report a rare case of very late dislodgement of a lumenless pacing lead.
- To highlight potential long-term risks associated with specific lead designs.
- To emphasize the importance of continued surveillance in cardiac pacing.
Main Methods:
- Case report of an 88-year-old patient with permanent atrial fibrillation and bifascicular block.
- Successful implantation of a Medtronic SelectSecure 3830 lumenless lead for VVIR pacing.
- Follow-up over three years, followed by investigation of pre-syncopal episodes and device malfunction.
Main Results:
- The patient experienced pre-syncopal episodes three years post-implantation.
- Device interrogation revealed capture failure, elevated pacing threshold, and decreased impedance.
- Fluoroscopy confirmed macro-dislodgement of the lumenless lead, which had coiled in the pacemaker pocket.
Conclusions:
- Late lead dislodgement can occur with lumenless leads years after initial successful implantation.
- This case necessitates continued vigilance during long-term follow-up of patients with cardiac physiological pacing systems.
- Successful lead extraction and replacement with a conventional lead were performed.
Related Concept Videos
Hypothesis: Accept or Fail to Reject?
29.5K
The outcome of any hypothesis testing leads to rejecting or not rejecting the null hypothesis. This decision is taken based on the analysis of the data, an appropriate test statistic, an appropriate confidence level, the critical values, and P-values. However, when the evidence suggests that the null hypothesis cannot be rejected, is it right to say, 'Accept' the null hypothesis?
There are two ways to indicate that the null hypothesis is not rejected. 'Accept' the null...
There are two ways to indicate that the null hypothesis is not rejected. 'Accept' the null...
29.5K
Nuclear Stability
23.3K
Protons and neutrons, collectively called nucleons, are packed together tightly in a nucleus. With a radius of about 10−15 meters, a nucleus is quite small compared to the radius of the entire atom, which is about 10−10 meters. Nuclei are extremely dense compared to bulk matter, averaging 1.8 × 1014 grams per cubic centimeter. If the earth’s density were equal to the average nuclear density, the earth’s radius would be only about 200 meters.
To hold positively charged protons together...
To hold positively charged protons together...
23.3K
RNA Stability
35.7K
Intact DNA strands can be found in fossils, while scientists sometimes struggle to keep RNA intact under laboratory conditions. The structural variations between RNA and DNA underlie the differences in their stability and longevity. Because DNA is double-stranded, it is inherently more stable. The single-stranded structure of RNA is less stable but also more flexible and can form weak internal bonds. Additionally, most RNAs in the cell are relatively short, while DNA can be up to 250 million...
35.7K
Stability
418
The time response of a linear time-invariant (LTI) system can be divided into transient and steady-state responses. The transient response represents the system's initial reaction to a change in input and diminishes to zero over time. In contrast, the steady-state response is the behavior that persists after the transient effects have faded.
The stability of an LTI system is determined by the roots of its characteristic equation, known as poles. A system is stable if it produces a bounded...
The stability of an LTI system is determined by the roots of its characteristic equation, known as poles. A system is stable if it produces a bounded...
418
Phase-lead and Phase-lag Controllers
550
Understanding the working function of different types of controllers can be illustrated with practical analogies, such as adjusting a stereo's volume equalizer. Cranking up the bass involves a phase-lead controller, which functions as a high-pass filter, while increasing the treble uses a phase-lag controller, which acts as a low-pass filter. PD controllers, similar to high-pass filters, enhance the system's response to high-frequency components. PI controllers, akin to low-pass...
550
Stability of structures
523
In mechanical engineering, the stability of systems under various forces is critical for designing durable and efficient structures. One fundamental way to explore these concepts is by analyzing systems like two rods connected at a pivot point, O, with a torsional spring of spring constant k at the pivot point. This system is similar in appearance to a scissor jack used to change tires on a car. In this case, the arms of the linkage (equivalent to the rods in this system) are entirely vertical,...
523