Quantitative in vivo Cherenkov luminescence imaging and dosimetry of 86Y-NM600
Campbell D Haasch1, Malick Bio Idrissou1, Sydney Jupitz1
1Department of Medical Physics, University of Wisconsin School of Medicine and Public Health, University of WI-Madison, 1111 Highland Avenue, 7109, Madison, WI, 53705, USA.
Purpose:
The rapid expansion of radiopharmaceutical therapy (RPT) development demands scalable preclinical dosimetry methods. While PET and SPECT remain the gold standards, their low throughput and high cost limit large-cohort studies. Cherenkov luminescence imaging (CLI) offers a high-throughput alternative but suffers from depth-dependent attenuation and photon scatter that compromise quantitative accuracy. This work develops and validates a quantitative CLI methodology incorporating attenuation and scatter corrections to enable accurate preclinical dosimetry.
Methods:
Depth-dependent attenuation was characterized using a tissue-mimicking phantom to derive calibration coefficients. Photon scatter was modeled using GEANT4-generated Cherenkov spread functions (CSFs), applied in a depth-weighted iterative Richardson-Lucy deconvolution/reconvolution framework. The method was evaluated in NU/NU mice (n = 4) bearing MC38 tumors after injection of 86Y-NM600, an isotope suitable for both PET and CLI. Liver and tumor activities were quantified at four timepoints using PET and the proposed CLI method. Voxelized Monte Carlo dosimetry was performed for both modalities.
Results:
CLI-PET activity quantification yielded mean errors of 15.4% (liver) and 10.3% (tumor) over the first three timepoints. Tumor absorbed doses from CLI-derived synthetic PET images (3.4 ± 0.3 Gy/MBq) were statistically indistinguishable from PET-based estimates (3.2 ± 0.2 Gy/MBq, p = 0.31). Discrepancies increased at late timepoints due to low activity and background auto-luminescence.
Conclusions:
With appropriate depth-dependent optical attenuation calibration and Monte Carlo-derived ionizing scatter correction, CLI can provide quantitative biodistribution and dosimetry estimates comparable to PET. This approach enables high-throughput, low-cost in vivo dosimetry, expanding the feasibility of large-scale preclinical RPT studies and supporting translational radiopharmaceutical development.
Related Concept Videos
Quantitative Analysis
In quantitative analysis, two key measurements are made: the sample quantity and a property proportional to the amount of the analyte (the substance being analyzed). This forms the basis of the...
Quantitative Aspects of Drug-Receptor Interaction
Imaging Studies VII: Vascular Imaging
Equivalence: In Vitro and In Vivo Bioequivalence
X-ray Imaging
Brain Imaging
These technologies include computerized axial tomography (CAT or CT scans), positron-emission tomography (PET scans), magnetic resonance imaging (MRI), functional magnetic resonance imaging (fMRI), and Transcranial Magnetic...