CT- DImQ: An open-access platform for image quality assessment of CT systems- application to CT values and noise
Ainur Kazhybekova1, Peter McHale2, J M den Harder3
1University College Dublin (UCD), School of Physics, Ireland.
Purpose:
To present an open access platform for CT image quality assessment combined with anthropomorphic phantoms. To promote the accessibility of anthropomorphic phantoms, including 3D-printed affordable ones, in regular QC analysis of CT images reconstructed with filtered back projection (FBP), iterative (IR) and Deep-Learning (DLR) reconstruction algorithms.
Methods:
The Python-based platform CT-DImQ enables the advanced evaluation of CT values distribution through histogram analysis and noise characterization (noise maps and noise power spectrum (NPS)). The interactive GUIs allow the user to perform the CT data analysis adapting the output to their own requirements. The use case for CT-DImQ was CT anthropomorphic phantom data (thorax phantom with 3D-printed TPU lung vessel trees and PA-12 nodules) obtained under clinical conditions and two dose levels reconstructed with FBP, IR and DLR.
Results:
Histograms showed in general an underestimation of CT-values for the 3D printed materials and a overestimation for Teflon (spine surrogate), in DLR images, while air and PMMA-soft tissue were the same as with IR and FBP. Noise maps showed a strong noise reduction in uniform phantom parts, and to a lesser extent in vessel and nodules edges. NPS showed noise reduction with increasing dose, especially with FBP and iterative reconstruction, and differences in the noise reduction frequency dependence among reconstruction methods, which was corroborated visually by the more uniform appearance of vessels and nodules in DLR images.
Conclusion:
CT-DImQ combined with anthropomorphic phantoms represents a step towards advanced image quality evaluation in CT, closer to the clinical reality with patients and able to evaluate advanced reconstruction methods trained with patient data.
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Radiological Investigation I: X-ray and CT
Muscles of the Thorax
The diaphragm is at the core of thoracic musculature, the primary muscle involved in breathing. This expansive, dome-shaped muscle marks the division between the thoracic and abdominal cavities. It...
Veins of Thorax
The azygos vein, positioned just right of the midline and anterior to the vertebral column, begins at the junction of the right ascending lumbar and subcostal veins, terminating in the superior vena cava. This vein drains blood from the right side of the thoracic wall, thoracic viscera, and posterior abdominal wall.
The...