Normothermic Negative Pressure Ventilation Ex Situ Lung Perfusion: Evaluation of Lung Function and Metabolism
Published on: February 14, 2022
Ex-PALM: a negative pressure ventilation system for air leak evaluation in ex vivo lungs
Mark Vartazarmian1,2, Alexandre Abid2, Rodin Chermat2
1Department of Mechanical Engineering, Institute of Biomedical Engineering, Polytechnique Montréal, Montreal, QC, Canada.
Introduction:
Prolonged air leaks (PAL) are considered to be one of the leading causes of postoperative complications following lung surgery. There are currently no clinically relevant methods for efficiently and systematically evaluating the underlying causes of PAL. Here, we introduce a new intuitive, physiologically-representative system for ex vivo negative pressure ventilation of lungs, equipped with PAL-oriented features.
Methods:
Reproducibility and system capabilities were assessed using a lung simulation model capable of controlling the effective area of the defect, then validated with ex vivo specimens.
Results:
Our system enables dynamic control of ventilation using either pressure (PCV) or volume (VCV) targets, with respective standard deviations of ±0.08 cm H2O and ±2.1 mL with moderate air leaks (<1,000 mL/min) and respective standard deviations of ±0.18 cm H2O and ±11 mL with severe air leaks (>1,000 mL/min). Additionally, leak quantification features proved comparable to that of the Thopaz+ (Medela Healthcare, Baar, Switzerland), a standard commercial digital thoracic drainage system, offering sufficient resolution to differentiate among clinically relevant air leaks. In the lower leak ranges (<400 mL/min) across all methods of evaluations, there were no significant differences between measured leak rates. For higher leak ranges, although there remained no significant differences between the Ex-PALM methods evaluated, the Thopaz + proved to systematically report lower leak rates values (Thopaz+ 420.0 ± 10.0 mL/min vs. PCV-derived 449.0 ± 19.9 mL/min, p < 0.05) and (Thopaz+ 1,200.0 + 0.0 mL/min vs. PCV-derived 1,239.7 ± 21.1 mL/min, p < 0.001). Unlike current systems, coughing was predictably replicated using peak pressure targets ranging from 100 to 300 cm H2O with a standard deviation of ±1.30 cm H2O from target. Our system allows extraction of biomechanical parameters at every breath, with theoretically expected pressures matching experimental measurements with a goodness fit value (R2) above 0.95 for the vast majority of breaths.
Discussion:
The Ex vivo Pulmonary Air Leak Model (Ex-PALM) provides a preclinical PAL testing platform with high translational potential and applications in studying biomechanical mechanisms of PAL and developing intraoperative mitigation strategies.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Lung Capacity
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Definition and Measurement of Pressure: Atmospheric Pressure, Barometer, and Manometer
Negative Regulator Molecules