Development of Visually Indistinguishable Acoustic Coupling Pads for Double-Blind Focused Ultrasound Neuromodulation
Samantha F Schafer1, Norman M Spivak2,3,4, Andrew A E D Bishay2
1School of Biomedical Engineering, Science and Health Systems, Drexel University, Philadelphia, PA.
Background:
Transcranial focused ultrasound (tFUS) is an emerging noninvasive neuromodulation modality with the ability to target deep brain structures with high spatial precision. Despite its promise, rigorous evaluation of its efficacy is limited by the absence of reliable, fully double-blind sham methodologies.
Objective:
To develop and validate a pair of visually and mechanically indistinguishable acoustic coupling pads that enable true double-blind tFUS neuromodulation studies by providing either efficient ultrasound transmission or robust ultrasound blocking without altering participant or operator experience.
Methods:
Two coupling pads were engineered: a transmitting pad designed to allow <5% pressure amplitude loss relative to free-water propagation, and a non-transmitting pad designed to attenuate ultrasound by ≥40 dB. Both pads used a Dragon Skin 10 NV silicone base and were identical in size, appearance, flexibility, and handling. The non-transmitting pad incorporated an encapsulated air-based blocking layer using an open-cell polyethylene foam insert. Acoustic performance was evaluated in a water tank using a 650 kHz BrainSonix transducer and a calibrated needle hydrophone. Sound speed of the silicone material was measured using pulse-echo techniques.
Results:
Twenty-three matched transmitting and non-transmitting pad pairs were fabricated and tested. Transmitting pads demonstrated a mean attenuation of -0.41 ± 0.53 dB, satisfying the design criterion of minimal acoustic loss. Non-transmitting pads demonstrated a mean attenuation of -48.61 ± 4.33 dB, exceeding the required -40 dB threshold for effective sham conditions. The Dragon Skin 10 NV substrate exhibited a sound speed of 964.72 m/s and produced <2 mm axial focal shift for standard pad thicknesses, with no measurable change in focal width. Both pad types were visually and tactually indistinguishable, could not be differentiated by experienced operators or participants, and maintained mechanical integrity after repeated cleaning.
Conclusion:
These acoustically engineered coupling pads provide a practical and validated solution for achieving true single- and double-blind conditions in tFUS neuromodulation studies. By preserving identical sensory and procedural experiences while enabling precise control over ultrasound transmission, this approach addresses a critical methodological gap in human ultrasound neuromodulation research.
Related Concept Videos
Blind Procedures
Blinding
Spin–Spin Coupling: Two-Bond Coupling (Geminal Coupling)
The central atom need not be NMR-active because its electrons are affected by the electron polarization of the spin-active atoms. However, spin information is transmitted less effectively than in one-bond coupling, and 2J values are usually weaker than 1J values. The energy of...
Spin–Spin Coupling: Three-Bond Coupling (Vicinal Coupling)
The extent of coupling depends on the C‑C bond length, the two H‑C‑C angles, any electron-withdrawing substituents, and the dihedral angle between the involved orbitals. The...
G-protein Coupled Receptors
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):