Evaluation and implementation of an independent Kilovoltage X-ray-based imaging platform for carbon ion radiotherapy
Yixiao Guo1,2,3, Zhiqiang Liu1,4, Qingzhen Zhu5
1Institute of Modern Physics, Chinese Academy of Sciences, Lanzhou, Gansu, China.
Background:
Image-guided particle therapy (IGPT) has significantly advanced in recent years, particularly in the context of proton therapy. However, imaging solutions for carbon-ion radiotherapy (C-ion RT) remain limited.
Purpose:
This study introduces sliding-gantry cone-beam computed tomography (CBCT) and dual-panel digital radiography (DR) systems, both mechanically independent of carbon-ion delivery nozzles. We aim to evaluate the image quality metrics and verify the positioning accuracy of the imaging systems.
Methods:
Image quality was evaluated in terms of spatial resolution, low contrast resolution, image uniformity, and effective imaging area using a multi-purpose imaging phantom, Catphan 700 phantom, and ImageJ software. The influences of planning computed tomography (CT) slice thicknesses (1-5 mm), radiation quality settings (90-130 kV), and registration algorithms (bony, grayscale, and fiducial marker registrations) on positioning accuracy were assessed using anthropomorphic head-neck and thoracoabdominal phantom images. The clinical feasibility of both systems was validated in 22 enrolled patients.
Results:
The CBCT exhibited a lower in-plane spatial resolution (2.50 line pairs per millimeter (lp/mm)) than DR (2.80 lp/mm). Spatial resolution of the CBCT system was measured at 0.90 lp/mm using the CTP 714 module of the Catphan 700 phantom. Both systems achieved a low contrast resolution of 2.30%. DR provided superior image uniformity (1.12%-1.40%) compared with CBCT (2.20%). The effective imaging areas were comparable between the CBCT and DR systems (99.30%-99.50%). Positioning accuracy varied with planning CT slice thicknesses, radiation quality settings, and registration algorithms, showing mean translation displacements ranging from 0.01 to 0.48 mm. CBCT achieved inter-fraction translational positioning errors within 2 mm in 42.3% (22/52) of fractions and rotational positioning errors within 2° in 80.8% (42/52) of fractions, and DR achieved 33.8% (24/71) and 73.2% (52/71), respectively.
Conclusion:
The developed CBCT and DR systems achieved superior image quality and sub-0.5 mm positioning accuracy. These findings support the clinical feasibility of integrating CBCT and DR imaging systems into the C-ion RT workflow.
Related Concept Videos
Ions as Acids and Bases
Salts are ionic compounds composed of cations and anions, either of which may be capable of undergoing an acid or base ionization reaction with water. Aqueous salt solutions, therefore, may be acidic, basic, or neutral, depending on the relative acid-base strengths of the salt’s constituent ions. For example, dissolving the ammonium chloride in water results in its dissociation, as described by the equation:
X-ray Imaging
The Carbon Cycle
X-ray Crystallography
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Introduction to Test of Independence
The test statistic for a test of independence is similar to that of a goodness-of-fit test: