Development and implementation of an MRI-only simulation, planning, and treatment workflow for prostate radiotherapy
Reza Reiazi1, Yao Ding1, Sarath Vijayan1
1Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
Purpose:
We evaluated the feasibility of a magnetic resonance (MR)-only simulation, planning, and treatment (MROSPT) workflow for prostate cancer patients using synthetic computed tomography (sCT) generated from magnetic resonance imaging (MRI) data. By validating sCT-based dose calculations, we aimed to streamline radiotherapy workflows, eliminate the need for CT simulation, and enable reliable clinical implementation of MR-based radiotherapy for MR-linac (MRL).
Methods:
We developed a comprehensive workflow encompassing the entire process from initial consultation to treatment delivery. After developing the workflow, a retrospective dosimetric validation study was performed on nine men with prostate cancer. They underwent CT and MRI simulations, and sCTs were generated from the MRI data. Contours and intensity-modulated radiation therapy treatment plans were created on the reference simulation CT (rCT) and transferred to sCTs for dose-calculation comparisons. Dosimetric accuracy was evaluated using gamma analysis (dose/distance; 2%/2mm). Bulk density sCTs (bCTs) were created by overriding organ density values with their mean (bulk) sCT-determined densities. bCT based on sCT allows treatment planning directly on MRI for MRL workflow efficiency.
Results:
Minimal non-bone Hounsfield units (HU)-value differences between rCT and sCT (5.5 ± 2.9 HU for prostate) demonstrated the reliability of the sCT generation process. Dosimetric comparisons between treatment plans (rCT vs. sCT, rCT vs. bCT) showed agreement within ± 2% in gamma analysis, confirming robust accuracy. The gamma index pass rate for rCT versus sCT and rCT versus bCT were consistently > 95% using 2%/2 mm criteria. A dry run of the entire simulation-to-treatment workflow was successfully completed.
Conclusion:
The MROSPT workflow using sCT is clinically feasible and dosimetrically accurate for prostate cancer patients. Dose calculations based on sCT demonstrated high dosimetric agreement with simulation CT, with no statistically significant differences across all evaluated metrics. These findings support the adoption of sCT‑based planning for prostate cancer radiotherapy and suggest its potential applicability in other anatomical regions especially in the pelvis. Integration of robust quality‑assurance processes and treatment‑delivery flexibility will further enhance its clinical utility.
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Role of Communication in the Nursing Process II: Planning and Implementation
Nursing Process for Patient and Caregiver Teaching II: Planning and Implementation
Nursing Implementation
The five steps to implementing effective nursing care include reassessing the patient, reviewing and revising the existing nursing care plan, organizing the resources and care delivery, anticipating and preventing complications, and implementing nursing interventions.
Sampling Plans
Random sampling is a method where each member of the population has an equal chance of being selected for the sample. It involves selecting individuals randomly, often using random number generators or lottery-type methods. For example, when analyzing the properties of a...
Synthetic Biology
Golden rice
Golden rice is a genetically modified...