Related Experiment Video
Updated: Feb 12, 2026

Quantifying the Relative Thickness of Conductive Ferromagnetic Materials Using Detector Coil-Based Pulsed Eddy Current Sensors
Published on: January 16, 2020
A Dynamic Shim Approach for Correcting Eddy Current Effects in Diffusion-Prepared MRI Acquisition Using a Multi-Coil
Congyu Liao1,2, Jason P Stockmann3, Zhitao Li2
1Department of Radiology and Biomedical Imaging, University of California, San Francisco, California, USA.
A novel dynamic B0 shimming method using an AC/DC shim array corrects eddy current phase errors in diffusion-prepared MRI. This enables high b-value imaging without signal-to-noise ratio loss from magnitude stabilizers.
Area of Science:
- Magnetic Resonance Imaging
- Biophysics
Background:
- Diffusion-prepared MRI is crucial for neuroimaging.
- Eddy currents from diffusion gradients cause phase errors, degrading image quality.
- Magnitude stabilizers are often used but reduce signal-to-noise ratio (SNR).
Purpose of the Study:
- To develop a dynamic B0 shimming approach for correcting eddy current-induced phase errors in diffusion-prepared MRI.
- To enable high b-value imaging without SNR loss.
- To eliminate the need for magnitude stabilizers.
Main Methods:
- A 46-channel AC/DC shim array and amplifier system were constructed.
- Spin echo prescans measured eddy current-induced phase differences.
- An optimization framework computed compensatory shim currents for multi-shot 3D diffusion-prepared acquisitions.
Main Results:
- The AC/DC shim array effectively corrected undesirable eddy current effects.
- Whole-brain, cardiac-gated, multi-shot 3D diffusion-prepared imaging was achieved without magnitude stabilizers.
- Full SNR was preserved, enabling reliable diffusion encoding at b values up to 2000 s/mm².
Conclusions:
- A new strategy using an AC/DC shim array compensates for eddy current-induced phase errors in diffusion-prepared MRI.
- This approach eliminates the need for magnitude stabilizers.
- Efficient, high-quality diffusion imaging with retained signal sensitivity is now possible.
More Related Videos
12:27Noninvasive In Vivo Small Animal MRI and MRS: Basic Experimental Procedures
Published on: October 20, 2009
11:11Effects of Transcranial Alternating Current Stimulation on the Primary Motor Cortex by Online Combined Approach with Transcranial Magnetic Stimulation
Published on: September 23, 2017
Related Concept Videos
Eddy Currents
Other major applications of eddy currents appear in metal detectors and the braking systems of trains and roller...
Diffusion
DC Generator
AC Sources
DC Battery