Lesser tuberosity osteotomy repair in stemless anatomic shoulder arthroplasty: a biomechanical analysis of repair
Mitchell S Kirkham1, Brittany Percin2, Kyle B Christy3
1Department of Biomedical Engineering, University of Utah, Salt Lake City, UT, USA; Department of Orthopaedics, University of Utah, Salt Lake City, UT, USA; School of Medicine, University of Utah, Salt Lake City, UT, USA.
Background:
Lesser tuberosity osteotomy (LTO) repair in total stemless shoulder arthroplasty has a high nonunion rate and thus presents a challenge to shoulder surgeons. Improved repair techniques may mitigate nonunion from excessive fragment motion or inadequate compression. The purpose of this study was to evaluate the biomechanical properties of 3 LTO repair techniques in the setting of stemless shoulder arthroplasty. Tensionable cortical button (Button) and suture anchor (Anchor) constructs were compared to a reference suture with a lateral plate construct (Suture).
Methods:
Forty human cadaveric shoulders from 20 matched pairs were dissected and the lesser tuberosity osteotomized. In each pair, the LTO was repaired with a Suture and lateral plate construct on one side and on the other side, either a tensionable cortical Button or suture Anchor construct was used. Sutures were passed through the humeral components of a stemless arthroplasty system during the procedures. All specimens were cycled 1,000 times from 10 to 100 N at 1 Hz using a custom subscapularis cryoclamp. Cyclic construct gapping was recorded at regular intervals with a digital video system. All specimens were then loaded to failure, and failure load, displacement, mode, and construct stiffness were recorded. Statistical analyses compared the Suture constructs to their paired Button or Anchor constructs.
Results:
Compared to their Suture pairs, the Button group displayed no differences in construct gapping (P ≥ .138), but the Anchor group displayed up to 50% increased gapping from cycles 1-400 (P ≤ .049). The Suture construct supported approximately 25% higher loads prior to failure when compared to their paired Button and Anchor group (P ≤ .014). There were no differences in failure stiffness between the Sutures and their paired Buttons or Anchors.
Conclusion:
In this controlled laboratory study, the decreased rate of initial construct gap formation and greater failure load of the suture construct suggests an environment for superior in vivo healing of the LTO as a result of decreased micromotion, with clinical implications still to be determined. Furthermore, the greater failure load in the suture construct could prevent catastrophic failure of the LTO in the delicate postoperative period. In the setting of total shoulder arthroplasty, an LTO subscapularis repair using a suture with a lateral plate construct provides a biomechanically superior repair to either a tensionable cortical button or suture anchor repair.
More Related Videos
Related Concept Videos
Mismatch Repair
Mismatch Repair
The Mutator Protein Family Plays a Key Role in DNA Mismatch Repair
The human genome has more than 3 billion base pairs of DNA per cell. Prior to cell division, that vast amount of genetic...
Overview of DNA Repair
Chemically...
Base Excision Repair
The first step of...
Nucleotide Excision Repair
Nucleotide Excision Repair
Cells are regularly exposed to mutagens—factors in the environment that can damage DNA and generate mutations. UV radiation is one of the most common mutagens and is estimated to introduce a significant number of changes in DNA. These include bends or kinks in the structure, which can block DNA replication or transcription. If these errors are not fixed, the damage can cause mutations, which in turn can result in cancer or disease depending on which sequences are...