Philipp Wallimann1, Philip Wild2, Petros Martirosian3
1Department of Radiation Oncology, University Hospital Zurich and University of Zurich, Rämistrasse 100, 8091, Zurich, CH, Switzerland. philipp.wallimann@usz.ch.
Objective:
Hybrid MR-Linac devices allow the acquisition of functional MR images during a regular radiotherapy treatment workflow, which could improve target delineation and response assessment. In this work, we investigate the feasibility of acquiring arterial spin labeling (ASL) perfusion images on a 0.35 T MR-Linac.
Materials And Methods:
A signal model for the flow-sensitive alternating inversion recovery (FAIR) balanced steady-state free precession (bSSFP) ASL sequence was introduced and an optimized set of sequence parameters for 0.35 T was determined. The signal dependence on the sequence parameters using the MR-Linac was examined in the brain of two healthy volunteers. Repeatability of cerebral blood flow (CBF) measurements with an adjusted MR-Linac protocol were compared to ASL images from a clinical 3 T scanner in six healthy volunteers that were scanned twice on each device.
Results:
The signal model described the observed measurements well, except for the dependence on inversion time. CBF values in gray and white matter in the six volunteers were systematically higher than at the 3 T scanner, but they had a similar percentage repeatability coefficient of 16-17%.
Conclusion:
An adjusted 10-min FAIR bSSFP ASL protocol on a 0.35 T MR-Linac yielded repeatable CBF measurements. (glinicaltrials.gov NCT05565326, submitted 2022-09-29).
More Related Videos
05:53Labeling of Blood Vessels in the Teleost Brain and Pituitary Using Cardiac Perfusion with a DiI-fixative
Published on: June 13, 2019
05:23Author Spotlight: Noninvasive Cerebral Blood Flow Determination in Human Functional Brain Region for Diagnosis of Neurological Disorders
Published on: May 31, 2024
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
NMR Spectroscopy: Spin–Spin Coupling
Spin–Spin Coupling: One-Bond Coupling
Spin–Spin Coupling Constant: Overview
Qualitatively, any spin plus-half nucleus polarizes the spins of its electrons to the minus-half state. Consequently, the paired electron in the hydrogen–carbon bond must...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Spin–Spin Coupling: Two-Bond Coupling (Geminal Coupling)
The central atom need not be NMR-active because its electrons are affected by the electron polarization of the spin-active atoms. However, spin information is transmitted less effectively than in one-bond coupling, and 2J values are usually weaker than 1J values. The energy of...