Standardizing quantitative muscle MRI: an inter-vendor evaluation in healthy adults
Johanna Thomä1,2, Lionel Butry1,2, Martijn Froeling3
1Department of Neurology, BG-University Hospital Bergmannsheil, Ruhr-University Bochum, 44789, Bochum, Germany.
Quantitative muscle MRI (qMRI) showed good reproducibility in healthy subjects. However, significant differences between manufacturers necessitate corrections for multi-center studies.
Area of Science:
- Medical Imaging
- Neurology
- Biophysics
Background:
- Quantitative muscle MRI (qMRI) is crucial for detecting muscular alterations in neuromuscular diseases (NMDs).
- Multi-center studies are vital for rare NMD research, demanding comparable data across different MRI vendors.
- Manufacturer-related variability in qMRI parameters remains underexplored.
Purpose of the Study:
- To evaluate the inter-vendor reproducibility of qMRI parameters in healthy subjects.
- To assess manufacturer-related variability in qMRI measurements.
- To inform data harmonization for multi-center NMD studies.
Main Methods:
- Ten healthy volunteers underwent qMRI on Siemens and Philips 3T systems within one week.
- Analyzed twelve leg muscles using Dixon (fat fraction, FF), T2 mapping (water T2, wT2), and diffusion-weighted imaging (FA, MD, RD, λ1).
- Reproducibility assessed via ICC, coefficient of variation, t-tests, and Bland-Altman analysis.
Main Results:
- Most qMRI parameters demonstrated moderate to excellent reproducibility (e.g., ICCFF=0.923, ICCwT2=0.792).
- Axial diffusivity (λ1) showed poor reproducibility (ICC=0.361).
- Significant inter-vendor differences (p<0.05) were observed for all parameters, with biases from 3.6% (FA) to 44.2% (FF).
Conclusions:
- qMRI parameters are generally reproducible within healthy volunteers.
- Absolute qMRI values differ significantly between manufacturers, requiring correction for accurate data pooling.
- Addressing manufacturer-related discrepancies is essential for reliable multi-center and longitudinal NMD research.