Inter-test reproducibility of lung clearance index in adults with clinically stable CF and in healthy controls
Chantal Darquenne1, Janelle M Fine1, Ann R Elliott1
1Department of Medicine, University of California, San Diego, CA 92093-0623, United States.
Background:
- Multiple breath nitrogen washouts (N2 MBW) and its derived lung clearance index (LCI) are increasingly used in clinical applications, due in part to the availability of commercial equipment. However, there are limited longitudinal data describing inter-test reproducibility for tests performed in adults.
Methods:
- Nitrogen MBWs were performed in triplicate in 41 subjects with clinically stable CF and in 56 healthy controls with normal spirometry. Functional residual capacity (FRC), LCI and alveolar mixing efficiency (AME) measurements were obtained during two visits separated by 3-6 weeks using a commercial device.
Results:
- At both visits, LCI was significantly higher and AME significantly lower in the CF group than in the healthy controls. There was no significant difference in FRC between the two groups. The average actual difference in all three MBW parameters between visits were close to zero in both groups. The biological variability of LCI (AME) was ±15 % (±4 %) in controls and ±18 % (±9 %) in the CF group.
Conclusions:
- The reproducibility of LCI and AME between sessions separated by 3-6 weeks is similar in healthy and clinically stable CF adults. Our findings further suggest that, for the commercial device used in this study, a percentage change in LCI greater than ±15 % can be considered physiologically relevant and greater than the biological variability of the test.
Related Concept Videos
Hepatic Drug Clearance: Restrictive and Nonrestrictive Clearance
Most drugs undergo restrictive clearance, which is proportional to the...
Renal Clearance
Renal clearance refers to the volume of plasma cleared of a specific substance, such as creatinine, per unit of time. To measure clearance, urine samples are collected over a 24-hour period during each bladder voiding, followed by a single blood sample at the...
Lung Capacity
Drug Clearance: Overview
Drug clearance is not limited to renal excretion but encompasses all organs involved in drug elimination, including...
Clearance Models: Physiological Models
The organ's clearance rate depends on the blood flow to the organ and the extraction ratio (E). The extraction ratio describes the organ's...
Clearance Models: Compartment Models