Background:
In the PROACTIVE-HF (A Prospective, Multi-Center, Open Label, Single Arm Clinical Trial Evaluating the Safety and Efficacy of the Cordella Pulmonary Artery Sensor System in NYHA Class III Heart Failure Patients) trial, remote heart failure (HF) management using seated mean pulmonary artery pressure (mPAP) and vital signs was safe and resulted in a low rate of HF hospitalizations and mortality through 6 months.
Objectives:
The authors evaluated the effect of managing seated mPAP with the Cordella system on outcomes in patients with HF through 12 months.
Methods:
In a single-arm, open-label trial, conducted in 75 European and U.S. centers, the authors enrolled HF patients with NYHA functional class III symptoms, irrespective of ejection fraction, and recent HF hospitalization and/or elevated natriuretic peptides. The prespecified, powered, secondary effectiveness endpoint at 12 months required the HF hospitalization or all-cause mortality rate to be lower than a performance goal of 0.70 events/patient/12 months, established from previous hemodynamic monitoring trials. Device/system-related complications, pressure sensor failure, and serious adverse events were examined.
Results:
Between February 7, 2020, and March 31, 2023, 456 patients were implanted in a modified intent-to-treat cohort. The 12-month event rate was 0.36 (95% CI: 0.31-0.42), which was significantly lower than the performance goal (0.36 vs 0.70; P < 0.0001). There were no device/system-related complications or pressure sensor failures beyond the 6-month primary results (0.8% and 0.2%, respectively).
Conclusions:
Remote HF management using seated mPAP and vital signs in NYHA functional class III patients is safe and results in a low rate of HF hospitalizations and mortality over 12 months. These results support the use of seated mPAP monitoring and extend the evidence that pulmonary artery pressure-guided management improves HF outcomes. (A Prospective, Multi-Center, Open Label, Single Arm Clinical Trial Evaluating the Safety and Efficacy of the Cordella Pulmonary Artery Sensor System in NYHA Class III Heart Failure Patients [PROACTIVE-HF Trial]; NCT04089059).
More Related Videos
09:40Induction of Right Ventricular Failure by Pulmonary Artery Constriction and Evaluation of Right Ventricular Function in Mice
Published on: May 13, 2019
07:02A Large Animal Model for Pulmonary Hypertension and Right Ventricular Failure: Left Pulmonary Artery Ligation and Progressive Main Pulmonary Artery Banding in Sheep
Published on: July 15, 2021
Related Concept Videos
Heart Failure V: Medical Management
Treatment for Pulmonary Arterial Hypertension: Oxygen Therapy for Respiratory Failure
Oxygen therapy is vital in increasing and maintaining blood oxygen levels in PAH patients. As a result, it aids in reducing fatigue,...
Heart Failure II: Pathophysiology
Pathophysiology of Heart Failure
Heart Failure I: Introduction
Heart Failure VI: Adjunct Therapies