Growing construct forces in early-onset scoliosis: How do TGR and VEPTR compare?
Jacob Jordan1,2, Vincent Ruggieri2, Ben Sinder2
1Perelman School of Medicine, University of Pennsylvania, Philadelphia, PA, USA.
Introduction:
Growing construct expansion surgeries in the setting of early-onset scoliosis (EOS) involve surgeon's feel to determine the optimal amount of distraction force to maximize correction and spinal length gained without compromising anchor integrity. Growing construct designs may require different amounts of distraction force to lengthen the spine or chest wall depending on the device used and various patient factors. Understanding these variations could guide appropriate force levels for the rational design for next-generation remote/external distraction devices and ultimately patient/deformity-specific lengthening forces.
Aim/Objectives:
The purpose of this study was to determine distraction force levels for expansion events in traditional growing rods (TGRs) and vertical expandable prosthetic titanium rib (VEPTR) constructs across patients with various anchor attachment points and etiologies of EOS.
Methods:
With IRB approval, two different distraction devices (VEPTR distractor and TGR/rod distractor) were outfitted with electronic strain gauges to measure distraction force. Force data were recorded at a rate of 40 Hz during each surgery via a laptop connected to a PowerLab 8/35 data acquisition system. After each surgery, the distractor was validated on a custom calibration device. The data collected were then analyzed to determine the maximum force and average force during every distraction.
Results:
From February 2019 to June 2023, a total of 146 in vivo distractions were collected on 52 patients with VEPTRs and 28 in vivo distractions were collected on 9 patients with traditional growing rods. TGR procedures required significantly higher forces than VEPTR (437.89 N vs 349.10 N, p < 0.001).
Conclusion:
Modification of two different distraction devices with electronic strain gauges identified differences in distraction force between traditional growing rods and VEPTR constructs during expansion surgeries in children with EOS. Our findings highlight that implant type and potentially other factors may play significant roles in the amount of force required for lengthening the spine and chest wall. Current market implant technology (MCGR maximum force of 187 N) is likely underpowered compared to traditional technology across all construct types and disease classifications. These insights should inform the development of next-generation remote lengthening construct design.
Related Concept Videos
Comparing Intermolecular Forces: Melting Point, Boiling Point, and Miscibility
Temporary attractive forces like dispersion are present in all molecules, whether they are polar or nonpolar. They...
Intermolecular vs Intramolecular Forces
Intermolecular Forces
Electromotive Force
Intermolecular Forces in Solutions
When the strengths of the intermolecular forces of attraction between solute and solvent species in a solution are no different than those present in the separated components, the solution is formed with no accompanying energy change. Such a solution is called an ideal solution. A mixture of ideal gases (or gases such as helium and argon,...
Comparing Copy Number Variations and SNPs
Copy number variations or CNVs are the structural variations that cover more than 1kb of DNA sequence. The single nucleotide polymorphism (SNP), on the other hand, is a single nucleotide change or a point mutation that is found in more than 1%...