Analytical performance of the Roche Elecsys TnT hs Gen 6 immunoassay.
Marco Tosi1, Laura Pighi1,2, Claudia Lo Cascio2
1Section of Clinical Biochemistry, University of Verona, Verona, Italy.
Clinical Chemistry and Laboratory Medicine
|February 16, 2026
Summary
The new Roche Elecsys TnT hs Gen 6 (TNT6ST) immunoassay shows improved precision and sensitivity for detecting cardiac injury. It also demonstrates better robustness against hemolysis interference compared to the previous version.
Area of Science:
- Clinical Chemistry
- Cardiovascular Diagnostics
- Immunoassay Technology
Background:
- Acute coronary syndrome (ACS) is a major global health concern requiring rapid diagnosis.
- High-sensitivity cardiac troponin (hs-cTn) assays are crucial for identifying myocardial injury.
- Evaluating novel diagnostic tools is essential for improving patient outcomes.
Purpose of the Study:
- To assess the analytical performance of the new Roche Elecsys TnT hs Gen 6 (TNT6ST) immunoassay.
- To compare the TNT6ST assay with the previous Elecsys hs-cTnT immunoassay (TNTHSSTX).
- To evaluate the impact of hemolysis on the performance of both assays.
Main Methods:
- Assessed intra- and inter-assay imprecision, linearity, LoB, LoD, functional sensitivity, and 99th percentile URL for TNT6ST.
- Compared TNT6ST performance against TNTHSSTX.
- Investigated hemolysis interference in plasma samples for both assays.
Main Results:
- TNT6ST demonstrated excellent precision (CVs 0.8-2.4%), linearity (r=1.00), and sensitivity (LoD 1.201 ng/L, functional sensitivity 1.255 ng/L).
- The 99th percentile URL was established (overall 20.9 ng/L, males 30.5 ng/L, females 13.7 ng/L).
- TNT6ST showed strong correlation (r=0.995) with TNTHSSTX but overestimated cTnT by ~66%; it also exhibited superior robustness to hemolysis.
Conclusions:
- The Roche TNT6ST immunoassay offers superior precision, sensitivity, and hemolysis robustness compared to its predecessor.
- The assay demonstrated excellent correlation across a wide cardiac troponin T range.
- These findings support the reliability of the TNT6ST assay for routine clinical use in diagnosing myocardial injury.