Design Differences and Usability Risks in GLP-1 Pen Injectors: A Comparative Study of a Generic and Reference Device
Laurie Brunet-Manquat1, Anne Combedazou2, Claire Ramus1
1BD Medical-Pharmaceutical Systems, Becton Dickinson and Company, Le Pont-de-Claix, France.
A new generic glucagon-like peptide-1 (GLP-1) receptor agonist injector shows comparable usability to the reference device. This suggests potential substitutability for Abbreviated New Drug Application (ANDA) submissions without additional user training.
Area of Science:
- Pharmaceutical Sciences
- Biotechnology
- Medical Device Engineering
Background:
- Glucagon-like peptide-1 (GLP-1) receptor agonists are crucial for managing type 2 diabetes.
- The development of generic combination products requires rigorous evaluation of delivery devices.
- Ensuring seamless substitution of delivery devices is key for Abbreviated New Drug Application (ANDA) submissions.
Purpose of the Study:
- To assess the usability and potential substitutability of a proposed generic GLP-1 receptor agonist injector.
- To determine if the generic injector can replace the Reference Listed Drug (RLD) device without additional training.
Main Methods:
- A comparative Human Factors study involved 14 participants (patients and caregivers) experienced with the RLD.
- Participants simulated critical tasks using both the generic and RLD delivery devices.
- Data collected included use error rates, observational data, and subjective feedback on ease of use and acceptability.
Main Results:
- The generic injector had a slightly higher use error rate (79%) versus the RLD (71%), with most errors not device-specific.
- 93% of participants showed similar usability profiles across both devices.
- Design differences were minor, with errors quickly resolved, indicating a rapid learning effect; 87% were comfortable switching devices.
Conclusions:
- The proposed generic GLP-1 injector demonstrates comparable usability to the RLD pen, posing no additional risks.
- Observed use errors were minimal, quickly resolved, and likely to have limited clinical impact.
- Findings support the potential substitutability of the generic device for ANDA submissions, aiding regulatory de-risking.
More Related Videos
Related Concept Videos
Insulin Formulations: Types and Delivery
Short-acting insulins are divided into...
Formulation and Manufacturing Process: Physical Attributes of Generic Tablets and Capsules
Parenteral Drug Delivery Systems: Injectables, Implants, and Infusion Devices
Glucagon-like Receptor Agonists
GLP-1, when administered in high doses intravenously, triggers insulin secretion, inhibits glucagon release, slows gastric emptying, reduces food intake, and restores normal insulin secretion. However, its rapid inactivation by...
Oral Hypoglycemic Agents: Glinides
Dosage Regimen: Individualization