Accuracy and performance of the Yuwell Anytime CT3 continuous glucose monitoring device
Matthew Lancaster1, Hope Edwards2, Dawn Harper3
1Associate Professor in Exercise Physiology, School of Biomedical Sciences, University of Leeds.
Background:
There have been many recent advances in diabetes technology, and continuous glucose monitoring (CGM) systems have grown in use and popularity.
Aims:
This study evaluated the accuracy and efficacy of the Yuwell Anytime CT3 CGM system in participants with type 2 diabetes.
Methods:
Forty participants were recruited to take part. They were asked to monitor their glucose levels for 10 days using two devices, a continuous glucose sensor (Yuwell Anytime) and an Accu-chek blood glucose finger-prick test.
Findings:
The overall mean absolute relative difference (MARD) between the two measurement systems was 6.9%, with a Pearson correlation coefficient of 0.92 (P<0.0001). The reliability of the sensor was preserved across the full observed range of glucose concentrations.
Conclusions:
The Yuwell Anytime had excellent agreement with a standard finger-prick blood test of glucose in free-living type-2 diabetics.