Perimetry of the Central Visual Field Using a Head-Mounted Open-Source Perimeter in Patients with Inherited Retinal
Cord Huchzermeyer1,2, Friedrich Kruse1,2, Jan Kremers1,2
1Department of Ophthalmology, University Hospital Erlangen, 91054 Erlangen, Germany.
Head-mounted perimeters (HMPs) offer a cost-effective way to assess visual fields in inherited retinal diseases (IRDs). The Iowa-HMP shows promise for research, despite limitations in dynamic range.
Area of Science:
- Ophthalmology
- Medical Technology
- Visual Science
Background:
- Head-mounted perimeters (HMPs) utilize consumer electronics for affordable visual field testing.
- HMPs have not yet been validated for patients with inherited retinal diseases (IRDs).
Purpose of the Study:
- To pilot-test the Iowa-HMP system for visual field assessment in patients with IRDs.
- To compare the Iowa-HMP's performance against a standard Octopus 900 perimeter.
Main Methods:
- The Iowa-HMP, comprising a smartphone, headset, and open-source software, was used.
- The 10-2 grid and ZEST algorithm were employed with a 10 cd/m² background.
- Testing included one normal subject and patients with macular dystrophy, Stargardt's disease, and retinitis pigmentosa.
Main Results:
- Typical visual field loss patterns were discernible with the Iowa-HMP.
- The Iowa-HMP demonstrated a dynamic range of 14 to 30 dB sensitivity.
- Limits of Agreement were ±7.5 dB, with diagnostic sensitivity of 0.67 and specificity of 0.95 for detecting low sensitivity locations.
Conclusions:
- The Iowa-HMP shows potential for visual field testing in IRDs, particularly for research.
- Open-source software facilitates further investigation in this field.
- Direct comparison with standard perimetry is limited due to the HMP's dynamic range.
More Related Videos
07:45Assessing Binocular Central Visual Field and Binocular Eye Movements in a Dichoptic Viewing Condition
Published on: July 21, 2020
07:12Development of a Gaze-Contingent Display Framework Designed for Perceptual and Oculomotor Research with Simulated Central Vision Loss
Published on: April 11, 2025
Related Concept Videos
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Angle Closure Glaucoma: Treatment