The Post-Marketing Real-World Outpatient Clinical Validation of Electronic Blood Pressure Monitors
Zhiping Zhang1, Muge Fan2, Qingsong Zhao1
1Shanghai General Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, People's Republic of China.
Abstract:
This study evaluated the accuracy of a variety of home upper-arm oscillometric blood pressure monitors (BPMs) from 448 participants against the FDA-certified XYZ110 auscultatory device using simultaneous measurements. Descriptive statistics were used to summarize the key findings of the study. Device performance was evaluated referencing international standards (ISO 81060-2:2013), where a mean absolute difference of ≤5 mmHg was defined as "Accurate." In addition, survey questionnaires completed by a subset of participants were analyzed to provide supplementary insights. A total of 448 consecutive outpatient clinical patients attending a routine clinical visit with their BPMs (male 215 and female 233) were eligible to participate in this study. The overall mean age of the participants was 62.35±12.59 years. Most of the BPMs included in this study were of the Omron brand (79.69%, n = 357), followed by Yuwell (5.8%, n = 26) and others. 76.32% of systolic blood pressure (SBP) measurements and 69.89% of diastolic blood pressure (DBP) measurements from BPMs exhibited differences of ≤5 mmHg compared to the reference calibrator. 70.89% (n = 318) were accurate in measuring SBP, and 60.27% (n = 270) were accurate in measuring DBP. Our study revealed that over 20% of BPMs exhibited discrepancies of more than 5 mmHg compared to a reference calibrator. Overall, the Omron U30, HEM-7211, and U10 models demonstrated relatively higher accuracy in blood pressure measurements. Regular patient to patient validation of BPMs is crucial to ensure accurate measurements for daily use.
More Related Videos
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
14:52Computerized Dynamic Posturography for Postural Control Assessment in Patients with Intermittent Claudication
Published on: December 11, 2013
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Measurement of Blood Pressure
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: