Experimental determination of effective X-ray attenuation coefficients of 3D-printed materials under clinical
Adrián Belarra1, Irene Hernández-Girón2, Julia Garayoa3
1Medical Physics, Radiology and Rehabilitation Department, Facultad de Medicina, Universidad Complutense de Madrid, Madrid, Spain.
Background:
3D printing enables the fabrication of customized breast phantoms for image quality assessment in digital mammography (DM) and digital breast tomosynthesis (DBT). A major challenge is the absence of standardized, accessible methods to characterize the attenuation properties of 3D-printed materials under clinical DM/DBT spectra.
Methods:
An experimental framework was implemented to determine the effective X-ray attenuation coefficient (μ eff ) of six 3D-printed polymers (PLA, PET, resin, ABS, ABS+, HIPS) and reference breast tissue-equivalent materials (CIRS plates simulating different breast glandular/adipose ratios (BR) and PMMA) using two commercial DM/DBT systems, with and without anti-scatter grid. Step-wedges (0.5-5.5 cm) were imaged across multiple kVp and filter settings. The μ eff were obtained from measurements on images and fitted to an empirical model yielding μ 0 (attenuation at thickness tending to zero) and k (decay rate) to characterize beam hardening and scatter influences. 3D-reference material equivalences were evaluated based on μ eff and μ 0 .
Results:
Beam hardening and scatter reduced μ eff with thickness, by 6%-14% with grid and 12%-28% without grid, with scatter contributing 47%-76% of the reduction in no-grid acquisitions. No significant differences were observed between the two mammography systems. Based on μ eff values, attenuation equivalences (within ±6%) were identified between 3D-printed and reference breast tissue-equivalent materials: PLA with BR 100/0; PET and resin with BR 70/30 and PMMA; ABS+ with BR 30/70 and BR 50/50. ABS and HIPS showed larger mismatches. The empirical model achieved excellent fits (R2 > 0.99), with μ 0 values preserving attenuation ranking and enabling derivation of equivalent glandular proportions.
Conclusion:
This framework demonstrates that routine clinical mammography systems can be used directly, without specialized instrumentation, to characterize 3D-printed materials as tissue surrogates. Several low-cost, widely available polymers were shown to reproduce breast tissue attenuation, supporting the local fabrication of anthropomorphic breast phantoms for realistic and clinically relevant image quality evaluation.
More Related Videos
Related Concept Videos
X-ray Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...