Related Experiment Video
Updated: May 12, 2026

Precision of In Vivo Quantitative Tooth Wear Measurement Using Intra-Oral Scans
Published on: July 12, 2022
Effect of Triton all-in-one irrigant on electronic working length determination using two apex locators: an in vitro
Aybüke Karaca Sakallı1, İsen Güleç Koçyiğit1, Bilge Özcan2
1Faculty of Dentistry, Department of Endodontics, Ankara Medipol University, Ankara, Turkey.
Background:
This study evaluated whether Triton, a newly introduced all-in-one irrigant, affects the accuracy of electronic working length determination, and compared two electronic apex locators (Ai-Pex, Propex Pixi) under different irrigant conditions.
Materials And Methods:
Forty-four extracted single-rooted human teeth were embedded in alginate. Actual working length was determined under a dental operating microscope. Specimens were assigned to four groups (Triton, 17% ethylenediaminetetraacetic acid, 10% citric acid, dry canal). Electronic working lengths were measured with Ai-Pex and Propex Pixi using #15 K-files. Statistical analyses were performed using one-way and two-way analysis of variance (ANOVA) following normality assessment, with statistical significance set at p < 0.05.
Results:
No significant differences occurred among Triton, ethylenediaminetetraacetic acid, and citric acid groups (p > 0.05). Dry canal measurements were significantly longer than the actual working length measured under a dental operating microscope (p < 0.05). Both apex locators showed comparable accuracy.
Conclusions:
Triton did not reduce apex locator precision, supporting its potential as a time-saving irrigant that permits simultaneous irrigation and measurement. Further clinical validation is required.