Universal Fiducial Markers for Multi-Modal Radiotherapy
Jie Liu1, Caroline Bamberger1, Naveed Iqbal1
1Inova Schar Cancer Institute, Inova Health System, Fairfax, VA, USA.
International Journal of Particle Therapy
|February 25, 2026
Summary
New coil-shaped fiducials improve visibility and tracking in radiotherapy while reducing proton beam perturbation. This universal design enhances clinical workflows across photon and proton therapy systems.
Area of Science:
- Medical Physics
- Radiotherapy Technology
- Biomedical Engineering
Background:
- Integrating photon and proton therapy requires universal fiducial markers for reliable performance across imaging and treatment systems.
- Standard fiducials face challenges: large sizes for computed tomography (CT) and image-guided radiotherapy (IGRT) cause proton beam perturbation, while smaller sizes compromise CyberKnife tracking accuracy.
Purpose of the Study:
- To design and implement customized fiducials that balance imaging visibility, tracking reliability, and minimal proton beam perturbation for multi-modality radiotherapy.
- To enable the use of universal fiducial markers across different radiotherapy systems, including photon and proton therapy.
Main Methods:
- Developed customized coil-shaped platinum fiducials (0.75 mm diameter) with variable coil gaps.
- Assessed visibility using cone-beam CT (CBCT) and kilovoltage (kV) imaging on photon and proton systems.
- Evaluated CyberKnife tracking reliability with an anthropomorphic phantom and respiratory motion platform.
- Measured proton beam perturbation using radiochromic films and a triple-channel film dosimetry method.
Main Results:
- Customized fiducials with increased coil gaps showed visibility comparable to standard fiducials in IGRT.
- These fiducials maintained reliable tracking performance in the CyberKnife system under static and dynamic conditions.
- Fiducials with coil gaps >0.5 mm/coil significantly reduced proton beam perturbation compared to standard fiducials.
Conclusions:
- Redesigned fiducial markers can maintain large diameters for visibility and trackability while minimizing proton beam perturbation by increasing coil gaps.
- A universal fiducial marker design is feasible, potentially streamlining clinical workflows and integrating various radiotherapy modalities.
Related Concept Videos
Positron Emission Tomography
Positron emission tomography (PET) is a medical imaging technique involving radiopharmaceuticals — substances that emit short-lived radiation. Although the first PET scanner was introduced in 1961, it took 15 more years before radiopharmaceuticals were combined with the technique and revolutionized its potential.
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Positron Emission Tomography (PET) is a medical imaging technique that provides crucial insights into the body's physiological functions at a molecular level. It is an indispensable resource for diagnosing, staging, and monitoring various illnesses, notably cancer, neurological disorders, and cardiovascular conditions.
Fundamental Principles of PET
Fundamental Principles of PET