Related Experiment Video
Updated: Feb 27, 2026

Multi-Modal Home Sleep Monitoring in Older Adults
Published on: January 26, 2019
In Vitro Investigation of the PneumoWave Biosensor for the Identification of Central Sleep Apnea in Pediatrics
Burcu Kolukisa Birgec1, Ross Langley2, Jennifer Miller3
1Strathclyde Institute of Pharmacy and Biomedical Sciences, University of Strathclyde, Glasgow G4 0RE, UK.
Abstract:
The interpretation and diagnosis of central sleep apnea in pediatrics by nocturnal polysomnography is challenging due to its technical complexity, which involves the simultaneous recording of multiple physiological parameters related to sleep and wakefulness. Furthermore, the unfamiliar environment of a sleep laboratory can hinder sleep evaluation, and diagnostic backlogs are common due to restricted capacity at specialist tertiary centers. The ability to undertake home sleep studies in a familiar environment using simple, robust, and low-cost technology is attractive. The potential to repurpose the PneumoWave biosensor, a UKCA Class 1 device, registered as an accelerometer-based monitoring device that is intended to capture and store chest motion data continuously over a period of time for retrospective analysis, was explored in an in vitro model of central sleep apnea. The PneumoWave system contains a biosensor (PW010), which was able to record simulated apnea episodes of 5 to 20 s across physiologically relevant pediatric breathing rates using an in vitro manikin model and manual annotation. The findings confirm that the PneumoWave biosensor could be a useful technology to support home sleep apnea testing and warrant further exploration.