Clinical Validation of rPPG-Enabled Contactless Pulse Rate Monitoring Software in Cardiovascular Disease Patients
Jing Wei Chin1, Po Him David Chan1, Shutao Chen1
1Department of Clinical AI, PanopticAI, Hong Kong SAR, China.
Insights
Remote photoplethysmography (rPPG) accurately measures pulse rate in cardiovascular disease (CVD) patients using video, offering a contactless solution for remote monitoring and early detection of health deterioration.
Area of Science:
- Biomedical Engineering
- Cardiology
- Medical Imaging
Background:
- Cardiovascular disease (CVD) is a leading global cause of death.
- There is a need for continuous, unobtrusive patient monitoring solutions.
- Remote photoplethysmography (rPPG) offers a contactless approach to vital sign measurement.
Purpose of the Study:
- To clinically validate the accuracy of rPPG for pulse rate (PR) measurement in patients with CVD.
- To compare rPPG-derived PR with traditional electrocardiogram (ECG) measurements.
- To assess the influence of various factors on rPPG accuracy.
Main Methods:
- 50 adults with confirmed CVD were enrolled.
- Synchronized facial video, ECG, and photoplethysmography (PPG) signals were recorded during a 6-minute session.
- PR was derived from 25-second video segments using rPPG software and compared to ECG-derived PR.
Main Results:
- rPPG-derived PR demonstrated strong agreement with ECG (Pearson correlation of 0.962).
- Mean absolute error was 1.061 bpm, with a root-mean-squared error of 2.845 bpm.
- Demographic, environmental, or CVD factors showed minimal influence on rPPG accuracy.
Conclusions:
- rPPG is an accurate contactless method for PR monitoring in CVD patients.
- This technology supports remote patient monitoring and early detection of deterioration.
- Further validation is needed for irregular rhythms, diverse populations, and additional vital signs.
Background:
Cardiovascular disease (CVD) is the leading cause of mortality worldwide, creating demand for continuous, unobtrusive monitoring solutions. This clinical validation evaluates the accuracy of remote photoplethysmography (rPPG), a contactless method using camera video, for measuring pulse rate (PR) in patients with CVD.
Methods:
We enrolled 50 adults with confirmed CVD at a clinical trial center. In a 6 min rested session, synchronized facial video (under controlled lighting), electrocardiogram (ECG), and photoplethysmography (PPG) signals were recorded. PR was derived from 25 s video segments using rPPG-enabled software and compared to ECG-derived PR via regression and Bland-Altman analysis.
Results:
Data from 47 participants (n = 817 samples) were analyzed. rPPG-derived PR showed strong agreement with ECG, with a mean absolute error of 1.061 bpm, root-mean-squared error of 2.845 bpm, and Pearson correlation of 0.962. Mixed-effects regression analyses (after 2% outlier removal, n = 782) indicated minimal influence from demographic, environmental, or CVD factors on accuracy. PPG-ECG discrepancies reflected inherent methodological differences.
Conclusion:
The rPPG method provides accurate, contactless PR monitoring in CVD patients, supporting its potential for remote patient monitoring and early deterioration detection. Future work will validate rPPG for irregular rhythms, additional vital signs, and diverse cohorts to strengthen clinical robustness for cardiometabolic risk assessment.
More Related Videos
06:16Signal Acquisition, Score Interpretation, and Economics of a Non-Invasive Point-of-Care Test for Coronary Artery Disease
Published on: August 9, 2024
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Special considerations while measuring pulse
Pulse
The pulse serves as a clinical...