Related Experiment Video
Updated: Feb 28, 2026

Dosimetry for Cell Irradiation using Orthovoltage 40-300 kV X-Ray Facilities
Published on: February 20, 2021
Quantifying uncertainty in calibration and clinical application of EBT-4 Gafchromic film dosimetry
Riccardo Lombardi1, Nicola Zancopè1, Samuele Cavinato2
1Department of Physics and Astronomy "Galileo Galilei", University of Padua, Padua, Italy.
Background:
GAFchromic films (GAFs) are essential tools in radiotherapy dosimetry, but uncertainties in calibration methods limit their accuracy and hinder consistency across clinical practice.
Purpose:
This study aimed to perform a comparative analysis of different calibration functions for GAFchromic EBT4 films in single- and double-channel dosimetry to quantify the uncertainty associated with their calibration and clinical use.
Methods:
GAFchromic EBT4 films were irradiated to doses between 0.2 and 10 Gy using a 6 MV photon beam. A custom irradiation setup in a water-equivalent phantom ensured reference conditions, with dose verified by an ionization chamber. Films were scanned on an Epson Expression 13000XL flatbed scanner at 72 dpi, with pre- and postirradiation scans taken at optimized time points (48 h postirradiation to ensure polymer growth stabilization). A custom-developed Python script was used for comprehensive film data processing, including lateral response artifact (LRA) correction. Eight calibration functions, including invertible and non-invertible forms from existing literature, were evaluated and compared. A comparison of the performance of the proposed approach with an established commercial software was performed.
Results:
Among the tested functions, two of them consistently emerged as the most accurate for EBT4 film calibration, yielding also the lowest relative errors. Specifically, the 2.5th order polynomial function showed an average percentage residual error of 2.7% (ranging from 1.6-4.4%), while the same function on corrected net optical density (netODc) demonstrated a comparable 2.9% (ranging from 1.9-4.4%). In general, calibration functions relying on netODc showed larger uncertainty compared to their analog using the red channel only. The best performing function showed residual values comparable to those obtained with the commercial benchmark. The LRA was observed to cause deviations of up to 5% at scanner extreme lateral positions, reinforcing the necessity of correction for large fields.
Conclusions:
We identified an optimal calibration function that incorporates a linear term and a 2.5th-order component to accurately model the relationship between radiation dose and the netODc in the red channel. The significant variability in performance among calibration functions underscores the critical need for independent verification and transparent dosimetry tools. To address this, our study makes the full analysis code and data publicly available, facilitating independent validation and enhancing the accuracy, consistency, and reproducibility of film dosimetry in clinical practice.
More Related Videos
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification ADCI and Dose Estimation
Published on: September 4, 2017
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Uncertainty in Measurement: Reading Instruments
Biological Effects of Radiation
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...