First evaluation of a heteroepitaxial diamond ionization chamber operating at low voltage for diagnostic X-ray
Kiyomitsu Shinsho1,2, Koji Koyama3, Keitaro Hitomi4
1Graduate School of Human Health Sciences, Tokyo Metropolitan University, Arakawa-ku, Tokyo, Japan.
Background:
Heteroepitaxial diamond has recently gained attention as a radiation detector material due to its wide bandgap, radiation hardness, and near-tissue equivalence. Despite these advantages, its use as a solid-state ionization chamber for diagnostic X-ray dosimetry has not yet been established. Demonstrating stable, high-efficiency operation at low voltage would enable compact dosimeters with a very small sensitive volume, which is difficult to achieve with conventional air ionization chambers.
Purpose:
To perform the first characterization of a heteroepitaxial diamond ionization chamber (HED-IC) operated at low bias voltage under diagnostic X-ray conditions and to evaluate its feasibility as a compact, high-efficiency dosimeter.
Methods:
A heteroepitaxial diamond detector (4 × 4 × 0.5 mm3) with Ti/Au electrodes was fabricated and evaluated using diagnostic X-ray beams at tube voltages from 50 to 120 kV. Charge-collection characteristics, dose linearity, energy dependence, and temporal response were assessed at negative bias voltages with magnitudes between -1 and -100 V. Monte Carlo simulations were performed using PHITS to compute the expected diamond-to-air sensitivity ratio under the same beam qualities for comparison with the experimental measurements.
Results:
The HED-IC exhibited excellent dose linearity (R2 > 0.997) and weak energy dependence (< 10%) across effective energies from 28.4 to 40.1 keV. The detector enables dose measurements within a very small sensitive volume, only 1/1250 of that of a typical air ionization chamber. The volume-normalized sensitivity exceeded theoretical expectations, suggesting enhanced effective ionization efficiency. An increased response with higher bias voltage further indicated potential for high-sensitivity operation.
Conclusions:
The results demonstrate that the HED-IC can operate as a low-voltage, high-efficiency solid-state ionization chamber under diagnostic X-ray conditions. Owing to the scalability of heteroepitaxial diamond growth, this detector concept provides a promising basis for compact, tissue-equivalent dosimeters capable of real-time dose monitoring across a wide range of radiological applications.
More Related Videos
06:51Dosimetry for Cell Irradiation using Orthovoltage 40-300 kV X-Ray Facilities
Published on: February 20, 2021
07:48An Externally-Heated Diamond Anvil Cell for Synthesis and Single-Crystal Elasticity Determination of Ice-VII at High Pressure-Temperature Conditions
Published on: June 18, 2020
Related Concept Videos
Determination of Crystal Structures
Atomic Emission Spectroscopy: Lab