Hydraulic endorectal actuator for prostate radiotherapy reduces variations in motion in a silicone rectal phantom
Aryan Niknam Maleki1, Mark Runciman1, Julia Murray2
1The Hamlyn Centre, Department of Surgery and Cancer, Imperial College London, London, United Kingdom.
Background:
The accuracy and morbidity of prostate cancer radiotherapy are influenced by unpredictable variations in rectal filling and patient motion. We developed a soft robotic hydraulic endorectal actuator that aims to reduce rectal motion and retract the rectum to restore the anorectal angle, improve target accuracy, and reduce toxicity during prostate cancer radiotherapy. The ability of the endorectal actuator to stabilize the rectum and improve prostate radiotherapy outcomes has not yet been assessed. This study evaluates the actuator's performance in a simulated rectal phantom.
Methods:
We fabricated a rectal phantom using silicone and motor-controlled elastic ribbons to simulate muscle tone and control the phantom diameter. The rectal compliance of the phantom was validated using a barostat balloon and was deliberately set low to simulate a high resistance to distension to challenge the actuator's capabilities. We assessed the actuator's ability to (1) resist dynamic peristaltic forces and (2) reproduce the rectal position and anorectal angle from varying initial displacements. The anterior-posterior rectal diameter and anterior rectal wall (ARW) displacements were measured using video tracker software.
Results:
The phantom demonstrated a rectal compliance of 4.19 ml/mmHg within the 40 ml-60 ml volume range, meeting the low-compliance target. During dynamic compression, the endorectal actuator reduced the change in the anterior-posterior diameter and ARW displacement from 25 mm and 15 mm, respectively, to less than 5 mm in both. The actuator reduced the increase in rectal volume from 132.3 cm3 (control) to 59.7 cm3 (actuator). When the phantom was translated anteriorly, the actuator reduced the anorectal angle deviation from +12° to +2° and anterior displacement of the ARW from 13 mm to 4 mm.
Conclusion:
Within this rectal phantom, the endorectal actuator reduced the variations in rectal motion. These findings suggest that the actuator may improve target accuracy and reduce radiation-induced toxicity in prostate radiotherapy, pending in vivo validation of the results.
Related Concept Videos
Drugs Affecting GI Tract Motility: Other Laxatives
Osmotic or saline laxatives, like magnesium hydroxide or milk of...
Drugs for Treatment of Constipation-Predominant IBS
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and solid...
Endoscopic Procedures V: ERCP
Patient...
Lower GI Series: Barium Enema
Procedure Details
The examination begins by inserting a lubricated rectal tube into the patient's rectum to administer a radiopaque barium solution. The barium flow is carefully...