Inappropriate defibrillator discharge despite evident atrial flutter on stored device electrogram having significant
Amit Varshney1, Amira Shaik1, Sanjeev S Mukherjee2
1Dept of Cardiology, RTIICS, Kolkata, India.
Analyzing device electrograms (EGMs) is key for therapy appropriateness. This case highlights inappropriate defibrillator therapy during atrial tachyarrhythmia due to misleading EGM analysis and algorithm deception.
Area of Science:
- Cardiology
- Biomedical Engineering
- Electrophysiology
Background:
- Device-stored electrograms (EGMs) are vital for assessing implantable cardioverter-defibrillator (ICD) therapy appropriateness.
- Inappropriate shocks can result from misinterpretation of atrial tachyarrhythmias as ventricular tachycardia (VT).