Inappropriate defibrillator discharge despite evident atrial flutter on stored device electrogram having significant
Amit Varshney1, Amira Shaik1, Sanjeev S Mukherjee2
1Dept of Cardiology, RTIICS, Kolkata, India.
Abstract:
It is crucial to analyze device stored electrograms (EGMs) to decipher whether the therapies were appropriate or inappropriate. We report a case of an inappropriate defibrillator therapy delivered during an obvious atrial tachyarrhythmia despite the dual chamber SVT discriminators being appropriately programmed for underlying rhythm abnormality. The therapy was delivered even though the tachycardia clearly satisfied V < A rate branch along with significant RR interval variability. The morphology match was misleading possibly due to malalignment of the peaks. On the other hand, the Abbott algorithm decides stability delta based on 2nd shortest and longest RR interval. This delta was <40 ms in spite of a irregular rhythm. The Atrio-Ventricular Association (AVA) algorithm, a second check-post designed for such situations, also got deceived due to variation of AV intervals. These led to fulfilment of stringent 'ALL' criteria programmed to make a diagnosis of VT and delivered therapy inappropriately.
Related Concept Videos
Mechanism of Cardiac Arrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Dysrhythmias III: Characteristics of Dysrhythmias
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Dysrhythmias VI: Management of Dysrhythmias