Delivered Oxygen Concentrations of Two Nonrebreather Reservoir Masks
Hatsuyo Ebisu1,2, Junko Takada3, Jhaymie Cappiello2
1Ms. Ebisu is affiliated with Tokyo Nishi Tokushukai Hospital, Akishima, Tokyo, Japan.
Background:
High-concentration nonrebreathing reservoir masks are frequently used for patients with hypoxemia but vary in shape, affecting mask fit. Mask fit affects the ability to inhale gas from the reservoir bag during inspiration and therefore makes it difficult to ensure a reliable concentration of oxygen delivered consistently to the patient. The EcoLite reservoir mask (EcoLite, Intersurgical, United Kingdom) reports a more consistent oxygen concentration delivery based on mask fit when compared with other products.
Methods:
A bench study was conducted comparing the consistency of high oxygen concentration delivery via a conventional reservoir mask (Inspiron, Japan Medicalnext, Japan) and EcoLite. The study was performed using an electronic breathing simulator and a mannequin to simulate spontaneous breathing. Each mask was applied to the mannequin per the product instructions. An oxygen analyzer was placed in the simulated left bronchus, and oxygen was delivered through each mask at increasing flow. The delivered oxygen concentration was recorded for each procedure and measured 3 times.
Results:
The EcoLite exhibited higher delivered oxygen concentrations at each oxygen flow and the simulator's inspiratory flow. The cumulative mean oxygen concentration across set parameters was 64% (SD ± 13) for the EcoLite and 50% (SD ± 10) for the Inspiron, with a significant difference (P < .008). Oxygen flow impact on delivered oxygen concentration revealed a difference between the EcoLite and Inspiron of P value > .05.
Conclusions:
EcoLite provided an oxygen concentration consistently higher than that of the conventional product observed. This may be attributed to mask fit, valve assembly, and reservoir bag function. These results may impact clinical decisions in product choice and applications. Further studies are needed to assess the clinical impact of this product.
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Cardiopulmonary Resuscitation II: ACLS Airway Management