Disposable Printed CamBlobs Charts for Measuring Contrast Sensitivity in Patients With Glaucoma.
Shahina Pardhan1, Rajiv Raman2, Rekha Srinivasan3
1Vision and Eye Research Institute, School of Medicine, Anglia Ruskin University, Young Street Cambridge, Cambridgeshire, CB1 2LZ, UK, anglia.ac.uk.
Journal of Ophthalmology
|March 2, 2026
Summary
The CamBlobs chart effectively assesses contrast sensitivity (CS) in glaucoma patients, showing comparable results to the Pelli-Robson chart. This simple, portable test offers a viable alternative for CS measurement in glaucoma care.
Area of Science:
- Ophthalmology
- Vision Science
Background:
- Contrast sensitivity (CS) is a crucial visual function often impaired in glaucoma patients.
- Established methods for CS assessment, like the Pelli-Robson chart, are valuable but may lack portability.
- There is a need for accessible and effective tools to measure CS in individuals with glaucoma.
Purpose of the Study:
- To evaluate the effectiveness of the inexpensive CamBlobs CS test in assessing contrast sensitivity in glaucoma patients.
- To compare the performance of the CamBlobs chart against the established Pelli-Robson CS chart.
- To determine the diagnostic accuracy and reliability of the CamBlobs test for glaucoma detection.
Main Methods:
- Two groups of participants (27 controls, 23 glaucoma patients) underwent CS testing using both CamBlobs and Pelli-Robson charts.
- Agreement and reliability were analyzed using Bland-Altman plots and receiver operating characteristic (ROC) curves.
- Visual field data were collected and correlated with CS measurements from both charts.
Main Results:
- Significant differences in CS were observed between control and glaucoma groups using both CamBlobs (p < 0.001) and Pelli-Robson (p < 0.001) charts.
- The CamBlobs chart demonstrated an area under the ROC curve (AUC) of 0.85, with 68.2% sensitivity and 85.2% specificity.
- The Pelli-Robson chart showed an AUC of 0.88, with 93.2% sensitivity and 66.7% specificity. Correlations with visual field mean deviation were r=0.59 for CamBlobs and r=0.65 for Pelli-Robson.
Conclusions:
- The CamBlobs CS test is comparable to the Pelli-Robson chart for assessing contrast sensitivity in individuals with glaucoma.
- CamBlobs provides a simple and portable method for CS testing, suitable for use outside traditional clinical settings.
- The CamBlobs chart is a valuable tool for evaluating visual function in glaucoma management.