Detailed Comparison of Westergren ESR and Automated 18°C ESR Using the YHLO VISION ESR Analyzer: A Comprehensive
Kartavya K Verma1, Girish S Kshatriya2, Himanshu Upadhyay3
1Pathology, Shri Rawatpura Sarkar Institute of Medical Sciences and Research, Raipur, IND.
Abstract:
Background Erythrocyte sedimentation rate (ESR) is a widely used, simple, and cost-effective test for diagnosing and monitoring conditions. Although the Westergren method is the gold standard, it has limitations like long processing times, infection risks, and a requirement for citrated blood samples. Automated ESR methods have been developed to overcome these issues. Materials and methods Blood samples were obtained from 4,000 patients who visited the central laboratory for ESR testing after providing informed consent. Each sample was first tested for ESR using the YHLO VISION ESR V4.0 Analyzer (Shenzhen YHLO Biotech Co., Ltd., Shenzhen, China) and subsequently analyzed with the manual Westergren method. The analysis included calculations of mean and standard deviation, Bland-Altman agreement assessment, linear regression, and Mann-Whitney U tests to evaluate the relationship between the two methods. Result Using data from 4,000 patient specimens, results demonstrated good agreement between methods with 94.65% of measurements within acceptable limits, a strong positive linear correlation (r = 0.9809, R² = 0.9621), and a systematic bias of 4.95 mm/h. Both methods effectively distinguished between patients with and without clinical evidence of elevated ESR (p < 0.001). Conclusions The 18°C measurement method provides a viable alternative with high clinical reliability and potential for cost-effectiveness.
Related Concept Videos
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
Raman Spectroscopy: Overview
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and...
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...