Cortisol assays across the physiologic spectrum: access immunoassay compared with LC-MS/MS.
Andrew J Newman1, Sanan Mahrokhian1, Isabelle Hanna1
1Center for Adrenal Disorders, Division of Endocrinology, Diabetes, and Metabolism, Mass General Brigham, Harvard Medical School, Boston, MA 02115, United States.
European Journal of Endocrinology
|March 5, 2026
Summary
This study demonstrates that the Access Cortisol immunoassay closely correlates with the gold-standard liquid chromatography-tandem mass spectrometry (LC-MS/MS) for measuring serum cortisol levels across a wide physiologic range.
Area of Science:
- Endocrinology
- Clinical Chemistry
- Assay Development
Background:
- Cortisol immunoassays are widely used for adrenal function testing.
- Previous studies have not comprehensively validated immunoassay performance against LC-MS/MS across the full spectrum of cortisol production.
Purpose of the Study:
- To compare the performance of the Access Cortisol immunoassay with LC-MS/MS for serum cortisol measurement.
- To assess immunoassay accuracy across morning, post-dexamethasone suppression, and post-cosyntropin stimulation conditions.
Main Methods:
- A cross-sectional study involving 97 adults without adrenal disease.
- Serum cortisol was measured using the Access Cortisol immunoassay and LC-MS/MS at three time points.
- Analysis included correlation and concordance assessments using common clinical cutoffs.
Main Results:
- The Access Cortisol immunoassay showed good correlation with LC-MS/MS across the physiologic spectrum.
- Agreement was closer at lower cortisol levels (post-dexamethasone suppression) than at higher levels (post-cosyntropin stimulation).
- High concordance (Cohen's κ > 0.89) was observed using standard interpretative cutoffs for all conditions.
Conclusions:
- The Access Cortisol immunoassay provides reliable cortisol measurements comparable to LC-MS/MS.
- Clinicians can be confident in using this immunoassay for evaluating adrenal function across varying cortisol levels.
Related Concept Videos
Mass Spectrometry: Complex Analysis
2.0K
Mass spectrometry is an important technique for the identification of pure compounds. However, it has some limitations for the analysis of complex mixtures, often due to excessive fragmentation making the spectrum too complicated to decipher. Mass spectrometry can be combined with suitable separation methods in sequence, forming hyphenated methods, which are useful in the analysis of complex mixtures.
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
2.0K
Enzyme-Linked Immunosorbent Assay
18.0K
In 1971, Peter Perlman and Eva Engvall developed an Enzyme-linked immunosorbent assay (ELISA or EIA). ELISA differs from western blot in that the assays are conducted in microtiter plates or in vivo rather than on an absorbent membrane.
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or...
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or...
18.0K