Inverted microscopy as a high performance tool for pre denudation evaluation of oocyte nuclear maturity
Nuri Peker1, Ayşen Yücetürk1, Özge Karaosmanoğlu1
1Acıbadem Maslak Hospital, In Vitro Fertilization Center, İstanbul, Türkiye.
Turkish Journal of Obstetrics and Gynecology
|March 5, 2026
Summary
An inverted microscope accurately predicts oocyte nuclear maturation before cumulus-oocyte complex (COC) denudation. This method offers high diagnostic accuracy, potentially improving in vitro maturation outcomes.
Area of Science:
- Reproductive Biology
- In Vitro Fertilization
- Oocyte Assessment
Background:
- Accurate oocyte nuclear maturation assessment is crucial for IVF success.
- Current methods lack definitive pre-denudation evaluation.
- Cumulus-oocyte complex (COC) communication is vital for maturation.
Purpose of the Study:
- To evaluate the inverted microscope as a reliable method for assessing oocyte nuclear maturation.
- To determine the diagnostic accuracy of inverted microscopy before COC denudation.
Main Methods:
- Prospective observational study involving 250 COCs from follicles <10 mm.
- Oocytes classified as germinal vesicle (GV) or non-GV using an inverted microscope.
- Diagnostic accuracy assessed after COC denudation and confirmation.
Main Results:
- Inverted microscopy achieved 96.91% accuracy for GV oocytes and 97.64% for non-GV oocytes.
- High concordance between pre- and post-denudation assessments.
- No significant difference in accuracy among embryologists.
Conclusions:
- Inverted microscopy provides a highly accurate method for predicting oocyte nuclear maturation.
- This technique can potentially enhance in vitro maturation by preserving COC communication.
- Enables informed decisions prior to oocyte denudation.
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
Phase-Contrast Microscopes
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Overview of Electron Microscopy
The wavelengths of visible light ultimately limit the maximum theoretical resolution of images created by light microscopes. Most light microscopes can only magnify 1000X, and a few can magnify up to 1500X. Electrons, like electromagnetic radiation, can behave like waves, but with wavelengths of 0.005 nm, they produce significantly greater resolution up to 0.05 nm as compared to 500 nm for visible light. An electron microscope (EM) can create a sharp image that is magnified up to 2,000,000X.
Overview of Microscopy Techniques
The early pioneers of microscopy opened a window into the invisible world of microorganisms. In 1830, Joseph Jackson Lister created an essentially modern light microscope. The 20th century saw the development of microscopes that leveraged nonvisible light, such as fluorescence microscopy that uses an ultraviolet light source and electron microscopy that uses short-wavelength electron beams. These advances significantly improved magnification, image resolution, and contrast. By comparison, the...