Intradialytic hypotension is severely underreported by intermittent blood pressure readings
Sabrine Chaara1,2, Muriel P C Grooteman1,2, Menso J Nubé1,2
1Department of Internal Medicine, division of Nephrology, Amsterdam University Medical Centre, Amsterdam, the Netherlands.
Background:
Intradialytic hypotension (IDH) has been associated with both morbidity and mortality. Yet, as IDH often occurs asymptomatic, it might be easily missed with routine twice per hour brachial artery cuff-blood pressure (BAC-BP) readings. This study evaluated differences in IDH detection between intermittent BAC-BP and continuous BP monitoring with a finger cuff (FC-BP). Validation of the FC device was an additional objective.
Methods:
In 40 chronic dialysis patients, intradialytic BP was measured during four sessions using (i) BAC-BP every 15 minutes and (ii) beat-to-beat FC-BP readings, which were averaged over 20-seconds intervals. IDH was defined as a systolic BP (SBP) <90 or <100 mmHg, depending on pre-dialysis SBP. Only FC measurements with an acceptable accuracy (i.e. mean SBP bias ≤10 mmHg) were used for the IDH analysis. For validation, sequential same-arm BP measurements were compared, using Bland-Altman analyses to assess agreement and four-quadrant plots to evaluate concordance.
Results:
For IDH analysis, 144 BAC-BP readings and 18 163 FC-BP measurements were recorded in 106 hours. While BAC-BP identified 20 IDH episodes, 441 were detected by FC-BP. For validation, Bland-Altman analyses of 1221 paired BP readings showed a mean bias ± limits of agreement of -13.3 ± 44.5 mmHg for SBP and -13.3 ± 25.5 mmHg for diastolic BP. Only 27.7% of the paired FC-BP readings met the acceptable accuracy criterion. Concordance rates were 34.3% and 36.3%, respectively.
Conclusions:
First, the occurrence of IDH is severely underestimated with intermittent BP monitoring. Second, FC-BP appears unsuitable for routine intradialytic BP monitoring due to its poor accuracy, precision, and concordance.
Related Concept Videos
Alterations in Blood Pressure
Hypertension (High blood pressure)
Hypertension occurs when blood pressure readings consistently exceed the normal range. It is diagnosed when systolic blood pressure (the top number, indicating pressure while the heart...
Errors occurring during blood pressure monitoring
Several factors...
Hemodialysis III: Nursing Management
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Pre-Procedural Guidelines for Assessing Blood Pressure