Method validation of cardiopulmonary exercise testing: intra- and inter-device comparisons across four metabolic
Robin Willixhofer1, Marlus Karsten1,2, Arianna Galotta1
1Centro Cardiologico Monzino IRCCS, University of Milan, Via Parea 4, Milan 20138, Italy.
Aims:
Cardiopulmonary exercise testing (CPET) is the gold standard for assessing exertional dyspnoea, providing peak oxygen uptake (VO2), and the VE/VCO2 slope. Different metabolic carts, may introduce systematic variability, confounding longitudinal assessments and multicentre trials.We evaluated intra-device reliability and inter-device comparability of CPET parameters.
Methods And Results:
In this prospective, single-centre, within-subject, repeated-measures, multi-device comparison study 34 healthy adults (35 ± 11 years; 59% males) completed eight CPETs: two per cart, on four systems (MGC Ultima™ CPX, COSMED Quark CPET™, Schiller PowerCube® Ergo, Vyaire Vyntus™ CPX). Tests used randomized order, ramp protocols on cycle ergometer, and standardized calibration, masks, and timing (2-10 days apart; within 40 days). Breath by breath data were 10-second averaged. Analyses included paired tests for intra-device reliability, repeated-measures ANOVA for inter-device comparisons, and iso-workload evaluation at 25-100% of lowest peak workload (ISO-Wpeak). Intra-device reliability was high, with no significant differences in VO2, heart rate or workload at anaerobic threshold (AT), respiratory compensation point (RCP) or peak. Minor variances occurred in resting VO2 (COSMED: 56 mL/min, P = 0.0427) and peak ventilation (Medical Graphic 9 L/min, P < 0.05). Inter-device comparability showed no differences in peak VO2 and the VE/VCO2 slope calculated up to the RCP. The VE/VCO2 slope assessed as the full slope however, showed significant differences between machines (P = 0.0048). Across standardized ISO-Wpeak (25%, 50%, 75%, and 100% of ISO-Wpeak), VO2 values were largely comparable between devices, with significant differences only for VO2 (mL/kg/min) at 25% ISO-Wpeak (P = 0.022).
Conclusion:
Under standardized conditions, the tested metabolic carts yield comparable values both for intra-device reliability and inter-device comparability.
Related Concept Videos
Exercise Stress Test
Exercise stress testing, commonly known as a treadmill test, is a noninvasive procedure used to evaluate cardiovascular function and diagnose heart conditions.
Definition
An exercise stress test measures the heart's response to exertion using a treadmill or stationary bicycle. Chest electrodes record the heart's electrical activity through an ECG, and blood pressure is monitored regularly.
Purposes
Pulmonary Function Tests
Pulmonary Function Tests are crucial diagnostic tools for assessing respiratory function, particularly in patients with chronic respiratory disorders. They comprehensively evaluate lung volumes, ventilatory function, breathing mechanics, diffusion, and gas exchange. These tests help diagnose pulmonary diseases and play a significant role in monitoring disease progression, evaluating disability, and assessing response to therapy.
PFTs involve using a spirometer, a...