Related Experiment Video
Updated: Jul 15, 2026

Treatment of Facial Deformities using 3D Planning and Printing of Patient-Specific Implants
Published on: May 23, 2020
Titanium mesh versus bioresorbable plates in orbital floor repair: Implications for blunt‑impact biomechanics
Minhee Hwang1, Shin Ae Lee2, Chi Hyun Lee3
1Department of Radiology, Pusan National University School of Medicine, Busan, Republic of Korea; Biomedical Research Institute, Pusan National University Hospital, Busan, Republic of Korea.
Background:
The native orbital floor behaves as a thin "blow‑out fuse," redirecting blunt‑impact energy away from the globe. Whether this protective role is preserved or lost after reconstruction depends largely on the stiffness and thickness of the implant, but quantitative evidence remains sparse.
Methods:
A subject‑specific, CT‑derived head model was analyzed using Abaqus/Explicit. A 20-mm cylindrical impactor struck the reconstructed orbit at 6 m/s. Three clinically used implants were compared: (i) 0.4 mm poly‑L‑lactide‑co‑glycolide / β‑TCP composite plate (CMP), (ii) 0.8 mm poly‑L‑lactide‑co‑glycolide plate (BMP), and (iii) 1.0 mm contoured titanium mesh (Titan fan). Mesh‑density and contact definitions were verified a priori. Peak von Mises stress, absolute maximum principal strain, and internal energy were extracted for the implant and globe.
Findings:
Peak globe stress/strain on the anterior surface was similar across reconstructions (von Mises stress 1.80-1.85 MPa). Implant behavior differed substantially: the titanium mesh reached a peak implant von Mises stress of 87.7 MPa while absorbing 14.2 mJ of internal energy. In contrast, CMP and BMP showed markedly lower peak implant stress (2.83 MPa and 2.75 MPa, respectively) while absorbing more internal energy (16.5 mJ and 44.0 mJ). Globe internal energy was 4332 mJ with titanium versus 4180 mJ with both bioresorbable plates (3.5% reduction).
Interpretation:
Compliant, bioresorbable plates dissipate a larger fraction of impact energy within the implant itself and modestly reduce the load transmitted to the eye. These results support retaining a degree of orbital‑floor compliance-rather than maximal rigidity-when selecting materials and thicknesses for reconstruction after blow‑out fracture.
Related Concept Videos
Cell-matrix's Response to Mechanical Forces
Anchoring junctions mechanically attach a cell to the...
Rolling Resistance
For instance, imagine a hard cylinder rolling on a comparatively soft surface. The cylinder's weight compresses the surface beneath it. As the cylinder moves, the material in front of it slows down due to...
Impact
When particles with different initial velocities collide, they induce deformation by applying equal and opposite impulses. At the point of maximum deformation, the particles move together with...
Types of Impact
The coefficient of restitution is a metric for understanding the dynamics of impacts. It quantifies the ratio of relative velocity...
Impact Loading
In cases of elastic deformation,...
Impact Loading on a Cantilever Beam
When an object is dropped onto the free end of a cantilever, its potential energy due to gravity is...