Gantry-based cone-beam CT of the thoracolumbar spine: A phantom comparison with Photon-Counting CT and
Ulf Bach1, Davide Cester1, Jonas Kroschke1
1Diagnostic and Interventional Radiology, University Hospital Zurich, University Zurich, Switzerland.
Background:
While cone beam CT (CBCT) is commonly used in musculoskeletal imaging of the extremities, its application in spinal imaging has been restricted by small field-of-view (FOV) coverage. Recent advancements in gantry-based CBCT systems promise to enable comprehensive imaging of the spinal column.
Purpose:
This study aimed to evaluate the performance of a novel gantry-based, multi-scan CBCT system for spinal imaging with complete anatomic coverage and compare it to energy integrating (EI)CT and photon counting (PC)CT using dose-matched protocols.
Materials And Methods:
An anthropomorphic torso phantom was used to simulate human anatomy. Gantry-based CBCT scans of the thoracolumbar spine were performed using different presets (low-dose, enhanced, best quality), while EICT and PCCT scans followed dose-matched clinical protocols. Qualitative image analysis was assessed by three blinded readers using a 4-point Likert scale, and quantitative analysis was conducted using global noise level (GNL) measurements.
Results:
CBCT achieved diagnostic-quality imaging for the thoracic and lumbar spine, particularly with "best" and "enhanced" presets. Subjective image quality was highest for PCCT, followed by EICT and CBCT. CBCT demonstrated lower GNL than EICT, nearing PCCT levels. However, high radiation doses (5 mGy) were required for CBCT imaging of the upper thoracic spine (Th1-Th6) due to anatomical complexity, while low doses (0,5 mGy) sufficed for the lower thoracolumbar spine (Th7-S1).
Conclusion:
Gantry-based CBCT was able to generate diagnostic-quality images of large spinal regions at relatively low radiation doses in a phantom setting, although the upper thoracic spine (above Th6) required higher doses. The overall subjective image quality remained below EICT and PCCT.
More Related Videos
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...