Ultrasound Assessment of the Third Ventricle Diameter: Agreement With CT Results in the Pediatric Population

Safae Dehbi1,2, Zhor Zeghari2,3, Chaimae Es-Sebbani4

  • 1Pediatric Intensive Care Unit (PICU) of the Children's Hospital of Rabat, IBN SINA University Hospital, Rabat, Morocco.

PubMed

Insights

Transcranial ultrasound (TUS) offers reliable bedside measurements of third ventricle diameter in children, showing good agreement with CT scans. This non-irradiating method is valuable for serial monitoring in pediatric intensive care units.

Area of Science:

  • Pediatric Neurology
  • Medical Imaging
  • Critical Care Medicine

Background:

  • Accurate measurement of the third ventricle diameter is crucial for managing critically ill children with neurological conditions.
  • Computed tomography (CT) is a standard imaging modality but involves radiation exposure.
  • Transcranial ultrasound (TUS) offers a non-irradiating alternative for bedside assessment.

Purpose of the Study:

  • To evaluate the agreement between transcranial ultrasound (TUS) and CT measurements of third ventricle diameter in critically ill children.
  • To determine the feasibility and reproducibility of TUS for ventricular assessment in a pediatric intensive care unit (PICU) setting.

Main Methods:

  • A prospective observational cohort study was conducted in a PICU.
  • 148 critically ill children (1 month to 15 years) underwent TUS within 1 hour of brain CT.
  • TUS measurements were performed using a phased-array probe through bilateral transtemporal windows.

Main Results:

  • TUS demonstrated high feasibility, with successful bilateral acoustic windows in 98.6% of cases.
  • The median third ventricle diameter measured 3.6 mm by both CT and TUS.
  • Bland-Altman analysis showed good agreement between TUS and CT, with minimal bias and acceptable limits of agreement.

Conclusions:

  • TUS provides reproducible bedside measurements of third ventricle diameter in children, comparable to CT in patients without midline shift.
  • TUS can serve as a valuable non-irradiating adjunct for serial ventricular monitoring, especially for hydrocephalus detection and external ventricular drain management.
  • Further research is recommended to assess interoperator reproducibility and TUS applicability in complex cases.