Backscatter from radiation protective drapes in fluoroscopically guided interventions - a patient dose to consider?
Pernilla Jonasson1, Charlotta Lundh1,2, Jens Hemmingsson2
1Department of Biomedical Engineering and Medical Physics, Sahlgrenska University Hospital, Gula Stråket 2B, Plan 0, Gothenburg SE-413 45, Sweden.
Radiation Protection Dosimetry
|March 13, 2026
Summary
Radiation protective drapes minimally increase patient radiation dose via backscatter. This minor increase is negligible when drapes are justified for occupational radiation protection during fluoroscopically guided procedures.
Area of Science:
- Medical Physics
- Radiological Protection
- Biomedical Engineering
Background:
- Radiation protective drapes are used in fluoroscopically guided procedures to shield medical personnel from scattered radiation.
- Concerns exist regarding potential increases in patient radiation dose due to backscattered radiation from these drapes.
Purpose of the Study:
- To quantify the backscattered radiation dose to patients from protective drapes.
- To assess the clinical significance of this backscattered radiation in fluoroscopically guided procedures.
Main Methods:
- Thermoluminescent dosimeters (TLDs) were used to measure backscattered radiation from three different protective drapes.
- Measurements were taken on top of and at various depths within a polymethyl methacrylate (PMMA) phantom.
- Monte Carlo simulations were employed to theoretically quantify backscattered radiation from lead materials.
Main Results:
- Backscattered radiation varied depending on the drape material.
- The measured backscattered radiation was less than 0.001 times the primary radiation beam dose.
- This included backscatter from the PMMA phantom itself.
Conclusions:
- The increase in patient radiation dose from protective drapes is minor and generally insignificant.
- The use of radiation protective drapes is acceptable if justified by occupational radiation protection benefits.
- Clinical decisions regarding drape use should weigh personnel safety against minimal patient dose increases.
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
566
The most common cardiovascular diagnostic test is an X-ray. It produces images of the heart, blood vessels, and adjacent structures.
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
566
Radiological Investigation I: X-ray and CT
1.5K
Radiological investigations, including X-rays and computed tomography (CT) scans, are critical for diagnosing and evaluating various medical conditions. These imaging techniques provide valuable insights into the body's internal structures, aiding in the detection of abnormalities, assessment of disease progression, and development of treatment strategies. This article delves into two primary radiological investigations, chest X-rays and CT scans, outlining their purpose, procedures, and...
1.5K