Agreement Between Two Swept-Source Optical Coherence Tomography Devices in Assessing Glistening on the Intraocular
Teresa Serrano González-Peramato1, José Ignacio Fernández-Vigo1,2,3, Beatriz De Pablo Gómez de Liaño4
1Department of Ophthalmology, Hospital Universitario Clínico San Carlos, Instituto de Investigación Sanitaria (IdISSC), 28040 Madrid, Spain.
Abstract:
Background/Objectives: To analyze the agreement between two swept-source optical coherence tomography (SS-OCT) devices in assessing glistening on intraocular lenses (IOL). Methods: Patients who had previously undergone cataract surgery were included. They were sequentially examined using two SS-OCT devices: Anterion (Heidelberg Engineering Inc., Heidelberg, Germany) and Triton (Topcon, Inc., Tokyo, Japan). Six corresponding scans from both devices were compared, and glistening, observed as hyperreflective foci (HRF), was manually counted. The total number of HRF and the degree of glistening were measured and categorized into four groups. The agreement between the two devices was analyzed using the intraclass correlation coefficient (ICC). Results: A total of 333 eyes from 285 patients were evaluated. The mean age was 76.5 ± 8.0 years (range: 45-95). The median number of HRF detected in a single scan was 1.1 (IQR 0.0-10.2, range 0-176) using Triton and 2.7 (IQR 0.2-20.1, range 0-250) using Anterion. The ICC across different scans ranged from 0.8 to 0.9, indicating strong agreement between the two devices. Bland-Altman plots showed better concordance in lenses with low glistening grades, while higher grades revealed greater discrepancies, with Anterion detecting significantly more HRF than Triton. Among all factors studied, only postsurgical time was associated with glistening. Conclusions: Two different SS-OCT devices can detect and quantify glistening in IOLs. The concordance between them was high, particularly for lower glistening grades. However, in higher grades, Anterion detected significantly more HRF than Triton.
More Related Videos
12:22Multimodal Volumetric Retinal Imaging by Oblique Scanning Laser Ophthalmoscopy oSLO and Optical Coherence Tomography OCT
Published on: August 4, 2018
06:55Scanning Light Scattering Profiler SLPS Based Methodology to Quantitatively Evaluate Forward and Backward Light Scattering from Intraocular Lenses
Published on: June 6, 2017
Related Concept Videos
Glaucoma: Overview
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...