Photon-Counting CT Enables Higher Image Quality at a Lower Radiation Dose Compared with a 256-Slice Energy-Integrating Detector CT in Pediatric Patients with Congenital Heart Disease
Related Experiment Video
Updated: Mar 15, 2026

High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
André Lollert1, Fabio Souschek1, Tariq Abu-Tair2
1Department of Diagnostic and Interventional Radiology, Section of Pediatric Radiology, Medical Center of the Johannes Gutenberg-University, 55131 Mainz, Germany.
Insights
Photon-counting CT (PC-CT) significantly reduces radiation dose and improves image quality in pediatric congenital heart disease patients compared to energy-integrating detector CT (EID-CT). This advancement offers safer, clearer imaging for this sensitive population.
Area of Science:
- Medical Imaging
- Radiology
- Pediatric Cardiology
Background:
- Pediatric patients with congenital heart disease require frequent cross-sectional imaging.
- This demographic is particularly vulnerable to ionizing radiation exposure.
- Existing CT technologies pose radiation risks that necessitate safer alternatives.
Purpose of the Study:
- To compare radiation dose and image quality between photon-counting CT (PC-CT) and energy-integrating detector CT (EID-CT).
- To evaluate the efficacy of first-generation PC-CT in pediatric congenital heart disease imaging.
- To determine if PC-CT offers a dose reduction benefit without compromising image quality.
Main Methods:
- Retrospective analysis of electrocardiography-gated thoracic CT scans in pediatric patients (<18 years).
- Comparison of radiation dose metrics (effective dose, CTDIvol, DLP) between PC-CT and EID-CT systems.
- Subjective (Likert scale) and quantitative (SNR, CNR) assessment of image quality by two independent observers.
Main Results:
- PC-CT group showed significantly lower effective dose, CTDIvol, and DLP compared to EID-CT.
- Subjective image quality and vascular contrast were rated higher in the PC-CT group.
- Quantitative contrast-to-noise ratio (CNR) was significantly superior in the PC-CT group.
Conclusions:
- First-generation PC-CT provides high-quality imaging for pediatric congenital heart disease evaluation.
- PC-CT achieves significant radiation dose reduction compared to conventional EID-CT.
- PC-CT represents a promising advancement for safer pediatric cardiac imaging.
Abstract:
Background/Objectives: There is an increasing need for cross-sectional imaging in pediatric patients with congenital heart disease. This patient group is highly sensitive to ionizing radiation. The purpose of this article was to evaluate differences in radiation dose and image quality between a first-generation photon-counting (PC)-CT system and a 256-slice single-source energy-integrating detector (EID)-CT in these patients. Methods: We retrospectively assessed effective dose, CT dose index-volume (CTDIvol), dose length product (DLP), and image quality of all prospectively electrocardiography-gated CTs of the thorax in all patients < 18 years of age examined between February 2021 and August 2024 (n = 43). Two independent observers subjectively scored image quality, vascular contrast, and noise on a 5-point Likert scale. In addition, we assessed the signal-to-noise-ratio (SNR) and contrast-to-noise-ratio (CNR) quantitatively. Results: All dose parameters were significantly lower in the PC-CT (n = 27) versus the EID-CT (n = 16) group (mean effective dose: 0.8 ± 0.64 versus 2.2 ± 0.88 mSv, p < 0.001; mean CTDIvol: 1.22 ± 0.96 versus 4.8 ± 1.08 mGy, p < 0.001; mean DLP: 30.7 ± 31.9 versus 73.7 ± 50.7 mGy*cm, p < 0.001). Overall subjective image quality and contrast were rated higher in the PC-CT group (p = 0.046 and < 0.001, respectively). Quantitative CNR was significantly higher in the PC-CT group (mean 39.1 ± 12.9 versus 26.2 ± 10.8, p = 0.002). Conclusions: PC-CT enables high-quality examinations for the evaluation of congenital heart disease with a highly significant dose reduction compared with a 256-slice single-source EID-CT.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation I: X-ray and CT