Wide-Field Oxygen Permeability Measurement of Contact Lenses Using a Modified Polarographic Electrode Cell
Wen-Hong Tong1, Jing Liu2, Jae-Yeon Pyo3
1Department of Optometry, Eulji University, 553 Sanseong-daero, Sujeong-gu, Seongnam-si 13135, Republic of Korea.
Sensors (Basel, Switzerland)
|March 14, 2026
Summary
A new wide-field method for measuring oxygen permeability (Dk) in contact lenses was developed. This improved technique provides more representative Dk values for better oxygen supply evaluation to the cornea.
Area of Science:
- Ophthalmic optics
- Biomaterials science
- Measurement science
Background:
- Oxygen permeability (Dk) is crucial for contact lens performance.
- Conventional Dk measurement methods have limited central measurement areas.
- This limitation may not reflect real-world oxygen transport under lens wear.
Purpose of the Study:
- To develop and evaluate a modified polarographic electrode cell for wide-field Dk measurement.
- To assess if expanded measurement area (11 mm diameter) improves representativeness.
- To ensure the new method meets ISO 18369 standards.
Main Methods:
- Developed a modified polarographic electrode cell for wide-field Dk measurement.
- Evaluated the system using plano hydrogel contact lenses of varying thicknesses.
- Compared accuracy, repeatability, and relative error against a conventional cell under ISO 18369 conditions.
Main Results:
- The modified cell achieved a wide-field measurement area of 11 mm.
- No statistically significant difference in Dk values was found compared to the conventional cell (p = 0.055).
- Relative error was 1.93%, meeting ISO acceptance criteria.
Conclusions:
- Wide-field Dk measurement is achievable without compromising reliability.
- The developed method offers a more representative evaluation of contact lens oxygen permeability.
- This approach is ISO-compliant and suitable for new testing method development.
Keywords:
ISO 18369contact lensescorneal oxygenationoxygen permeabilityoxygen transmissibilitypolarographic electrode cellwide-field measurementMore Related Videos
Related Concept Videos
Amperometry: Overview
2.0K
Amperometry is a technique commonly used to measure the concentration of specific analytes in a solution by monitoring the electric current generated during an electrochemical reaction. It involves applying a constant potential between a working electrode and a reference electrode to measure the resulting current, which is proportional to the concentration of the analyte. The Clark oxygen electrode operates based on this principle of amperometry. It consists of a cathode and an anode enclosed...
2.0K
Potentiometry: Membrane Electrodes
2.1K
Membrane electrodes, also known as p-ion electrodes, use membranes that selectively interact with free analyte ions, generating a potential difference across the membrane. The resulting membrane potential, known as the asymmetry potential, is not zero even when analyte concentrations on both sides of the membrane are equal. The membrane's response is typically not selective to a single analyte but proportional to the concentration of all ions in the sample solution capable of interacting at...